Dictionary · Uncertainty

Type A evaluation of measurement uncertainty

Evaluation of a component of uncertainty by statistical analysis of measured values obtained under defined conditions — e.g. the standard deviation of the mean of repeated readings.

Source
VIM (JCGM 200:2012) 2.28
Discipline
Uncertainty

Search the whole Atlas for “Type A evaluation of measurement uncertainty” →

Related terms

More from Uncertainty

Triangular distribution

A probability distribution rising linearly to a peak midway between limits ±a. Used when values near the centre are more likely; its standard uncertainty is a/√6.

GUM (JCGM 100:2008) 4.3.9

Coverage probability

The probability that the set of true quantity values of a measurand is contained within a specified coverage interval — commonly 95 % in calibration practice.

VIM (JCGM 200:2012) 2.37

Sensitivity coefficient

The partial derivative ∂f/∂xᵢ of the measurement model with respect to an input quantity, evaluated at the estimates: it converts an input standard uncertainty into its contribution to the uncertainty of the result.

GUM (JCGM 100:2008) 5.1.3

Type B evaluation of measurement uncertainty

Evaluation of a component of uncertainty by means other than statistical analysis of current measurements — calibration certificates, specifications, handbook data, assumed distributions.

VIM (JCGM 200:2012) 2.29

Definitional uncertainty

The component of uncertainty resulting from the finite detail in the definition of the measurand itself — the floor below which no better procedure can go without redefining the quantity.

VIM (JCGM 200:2012) 2.27

Uncertainty budget

A statement of a measurement uncertainty, its components, and their calculation and combination — usually tabulated with each source, evaluation type, distribution, standard uncertainty, sensitivity coefficient, and contribution.

VIM (JCGM 200:2012) 2.33