Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
GLP 10 – 2019 Page 1 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 GLP 10 Good Laboratory Practice for the Purity of Water 1 Water for volume calibrations. Water is u...
Open at page →
Page 2
GLP 10 – 2019 Page 2 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 salts are tolerable before density is affected, it is prudent to use the best system practical with...
Open at page →
Page 3
GLP 10 – 2019 Page 3 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 addition, D 1193 specifies that preventive maintenance shall be performed on a regular basis and no...
Open at page →
Page 4
GLP 10 – 2019 Page 4 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 Temperature equilibrium is a critical factor in the density stability of water. This is especially...
Open at page →
Page 5
GLP 10 – 2019 Page 5 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 2 999.97495 (1 (((G8 3.983035) (G8 301.797)) / (522528.9 (G8 69.34881))))= ×− − ×+ ×+ Eqn. (2) Air-...
Open at page →
Page 6
GLP 10 – 2019 Page 6 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 Example Data Table 2. Example data for water density calculations. Temperature (°C) Air-Free Water...