Page 1
GLP 9 – 2019 Page 1 of 4 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 GLP 9 Good Laboratory Practice for Rounding Expanded Uncertainties and Calibration Values A calibra...
Open at page →
Page 2
GLP 9 – 2019 Page 2 of 4 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 When the digit next beyond the one to be retained is less than five, keep the retained figure uncha...
Open at page →
Page 3
GLP 9 – 2019 Page 3 of 4 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 Example 2 The volume of a given flask is computed to be 2000.714 431 mL and the uncertainty is 0.08...
Open at page →
Page 4
GLP 9 – 2019 Page 4 of 4 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 Rather than stating the uncertainty value with each correction, it is also proper to place the corr...