GMP 10 – 2019 Page 1 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 GMP 10 Good Measurement Practice for Understanding Factors Affecting Weighing Operations Good labo...
Open at page →
Page 2
GMP 10 – 2019 Page 2 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 standards were stored in a cabinet and unknown test items near the balance where a temperature gra...
Open at page →
Page 3
GMP 10 – 2019 Page 3 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 Mechanical analytical balances are provided with partial and full release positions. The partial r...
Open at page →
Page 4
GMP 10 – 2019 Page 4 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 exercised. Balance drift is likely to be larger initially and then reach a steady state when the b...
Open at page →
Page 5
GMP 10 – 2019 Page 5 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 5 Comparison Weighing - Electronic Balance Measurements made on a full electronic balance are simp...
Open at page →
Page 6
GMP 10 – 2019 Page 6 of 6 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 201 9 charge on the balance. Electrostatic charges are of concern when plastic containers are placed on...