Page 1
SOP 17 – 2019 Page 1 of 5 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 SOP 17 Standard Operating Procedure for Control Charts of Check Standards 1 Introduction Purpose Th...
Open at page →
Page 2
SOP 17 – 2019 Page 2 of 5 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 its short-term precision. When the calibration process is determined to be in a state of statistica...
Open at page →
Page 3
SOP 17 – 2019 Page 3 of 5 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 Construct a Standard Deviation chart using the same approach. You may use 2 and 3 as the respective...
Open at page →
Page 4
SOP 17 – 2019 Page 4 of 5 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 Control the Measurement Process An appropriate check standard is calibrated each time the laborator...
Open at page →
Page 5
SOP 17 – 2019 Page 5 of 5 This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 - 2019 Appendix Check Standard Data for Control Chart Check Standard I dentification Nominal Value _______...
Open at page →
Page 6
This page is intentionally blank. This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383-20 19