Volume Transfer Graduated Provers
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 SOP 19 Standard Operating Procedure for Calibration of Graduated Neck-Type Metal Provers (Volume Transfer Method) 1 1 Introduc...
Open at page →
Page 2
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 international system of units (SI). Metrological traceability may be to the SI through a National Metrology Institute such as...
Open at page →
Page 3
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 2 Methodology Scope, Precision, Accuracy This procedure is applicable for the calibration of any size metal prover within the...
Open at page →
Page 4
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 reduce water retention errors. Care must be taken during wet-downs and runs to ensure complete drainage and consistent retenti...
Open at page →
Page 5
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 2.4.3.3 Run 1. Fill the standard and measure and record the temperature, t 1. 2.4.3.4 Adjust the standard prover to its refere...
Open at page →
Page 6
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 2.4.3.9 Perform the calculations described in section 3 to determine the prover volume at the appropriate reference temperatur...
Open at page →
Page 7
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Multiple Deliveries Calculate V X60, the volume of the unknown prover at 60 F, using the following equation: + ...
Open at page →
Page 8
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Prover Error/Correction or Deviation from Nominal The total calculated volume of the prover at its reference temperature shoul...
Open at page →
Page 9
Table 3. Common reference temperatures. This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Commodity Reference Temperature Frozen food labeled by volume (e.g., fruit juice) -18...
Open at page →
Page 10
Table 4 Variables for volume equations at alternative temperatures. This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Symbols Used in Equations V XtrefX volume of the unknown v...
Open at page →
Page 11
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 repeatability assessments and the source of the error cannot be determined conclusively and corrected, the measurement must be...
Open at page →
Page 12
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 A standard deviation chart is also used for measurement assurance through the evaluation of replicate measurements (See SOP 20...
Open at page →
Page 13
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 the process only and does not represent potential systematic differences in meniscus readings. In a laboratory with only one s...
Open at page →
Page 14
Table 5. Example Uncertainty Budget Table. This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Uncertainty Component Description Symbol Source Typical Distribution Uncertainty of...
Open at page →
Page 15
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Conformity Assessment Evaluate compliance to applicable tolerances as needed or required by the customer or by legal metrology...
Open at page →
Page 16
Appendix A Volume Transfer Data Sheet (page 1) Laboratory data and conditions: This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Vessel Owner Operator Vessel Identification Dat...
Open at page →
Page 17
This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Appendix A Volume Transfer Data Sheet (page 2) Run 1: Data from the filled unknown field standard: Material for the Unknown: M...
Open at page →
Page 18
Appendix B Example Temperature Correction Table A This publication is available free of charge from: https://doi.org/10.6028/NIST.IR.7383 -2019 Example for a 100 gallon prover and 60 °F reference temperature. Mild Steel...