SOP 33 – 2019 Page 1 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 SOP 33 Standard Operating Procedure for Calibration of Weight Carts 1 Introduction This Standard...
Open at page →
Page 2
SOP 33 – 2019 Page 2 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Standards. Ensure the availability of suitable working standards or recently calibrated (and unu...
Open at page →
Page 3
SOP 33 – 2019 Page 3 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Equipment /Apparatus/Assistance A balance, scale, or load cell with suitable resolution and repe...
Open at page →
Page 4
SOP 33 – 2019 Page 4 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 A basic knowledge of weight carts, including the braking system and operating controls is requir...
Open at page →
Page 5
SOP 33 – 2019 Page 5 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Weight carts that repeatedly demonstrate excessive “as found” out-of- tolerance conditions from...
Open at page →
Page 6
SOP 33 – 2019 Page 6 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 the largest and smallest differences. Record the range information in a suitable control chart f...
Open at page →
Page 7
SOP 33 – 2019 Page 7 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Tolerances for weight carts (See also NIST Handbook 105- 8). Tolerances are calculated as 0.035...
Open at page →
Page 8
SOP 33 – 2019 Page 8 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 standard deviation of the process is calculated each time a weight cart is calibrated. A minimum...
Open at page →
Page 9
SOP 33 – 2019 Page 9 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 10 Calibration Certificate Prepare a Calibration Certificate according to SOP 1. Report the mean...
Open at page →
Page 10
SOP 33 – 2019 Page 10 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Appendix A Sample Calibration Certificate for Weight Cart Test Number: Date Issued: Date of Cal...
Open at page →
Page 11
SOP 33 – 2019 Page 11 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 checklist may contribute measurement errors that are significant if not properly maintained dur...
Open at page →
Page 12
SOP 33 – 2019 Page 12 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Inspection Checklist The weight cart was cleaned and painted and allowed to come to thermal and...
Open at page →
Page 13
SOP 33 – 2019 Page 13 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Inspection Checklist for Weight Cart Test Number: Date of Inspection: Manufacturer: Date of Man...
Open at page →
Page 14
SOP 33 – 2019 Page 14 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 Appendix B Recommended Method for Intermediate Verification of Mass Values for Weight Cart User...
Open at page →
Page 15
SOP 33 – 2019 Page 15 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 2.3. Standards. Class F field standards are used if they are within the prescribed calibration...
Open at page →
Page 16
SOP 33 – 2019 Page 16 of 17 This publication is available free of charge from: https://doi.org /10.6028/NIST.IR .6969 - 2019 7. Tolerance and Evaluation The observed estimated error must be less than one third of the sma...
Open at page →
Page 17