Selection Use Sensitivity Weights Tare Weights
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
SOP 34 – 2019 Page 1 of 5 This publication is a vailable free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 2019 SOP 34 Standard Operating Procedure 1 for the Selection and Use of Sensitivity Weights and Tare We...
Open at page →
Page 2
SOP 34 – 2019 Page 2 of 5 This publication is a vailable free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 2019 Verify that the operator is familiar with the design and the operation of the balances and familia...
Open at page →
Page 3
SOP 34 – 2019 Page 3 of 5 This publication is a vailable free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 2019 Table 1. Recommended maximum difference between X and S . Balance (X – S) Equal arm balance each o...
Open at page →
Page 4
SOP 34 – 2019 Page 4 of 5 This publication is a vailable free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 2019 A sensitivity weight is not required if using an electronic mass comparator or fully electronic ba...
Open at page →
Page 5
SOP 34 – 2019 Page 5 of 5 This publication is a vailable free of charge from: https://doi.org/10.6028/NIST.IR.6969 - 2019 The applicable tolerance may be substituted for the average maximum difference between X and S, d....
Open at page →