Resource Library · SP 250
Liquid in Glass Thermometer Calibration Service
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
NISTMeasurementServices: Liquid-In-Glass ThermometerCalibration Service NIST Special Publication 250-23 JacquelynWise U.S.DepartmentofCommerce NationalInstituteofStandardsandTechnology
Open at page →
Page 2
CenterforBasicStandards TheCenterforBasicStandardsdevelopsandmaintainsthescientificcompetenciesandlaboratoryfacilitiesnec- essarytopreserveandcontinuetorefinethebasephysicalquantitiesuponwhichtheNation'smeasurement syste...
Open at page →
Page 3
NISTMEASUREMENTSERVICES: Liquid-In-GlassThermometerCalibrationService JacquelynWise TemperatureandPressureDivision CenterforBasicStandards NationalMeasurementLaboratory NationalInstituteofStandardsandTechnology Gaithersb...
Open at page →
Page 4
LibraryofCongressCatalogCardNumber:88-600580 NationalInstituteofStandardsandTechnology SpecialPublication250-23,128pages(Sept.1988) CODEN:XNBSAV U.S.GOVERNMENTPRINTINGOFFICE WASHINGTON: 1988 ForsalebytheSuperintendent of...
Open at page →
Page 5
PREFACE Calibrations andrelatedmeasurement servicesoftheNationalBureauof Standardsprovidethemeansformakersandusersofmeasuringtoolsto achieve levelsofmeasurement accuracy thatarenecessary toattain quality,productivity and...
Open at page →
Page 6
Thepresentpublication, SP250-23,NBSMeasurement Services:Liquid-in- GlassThermometer Calibration Service,byJacquelyn A.Wise, isoneof approximately 20documentsinthenewseriespublishedorinpreparationby theCenterforBasicStand...
Open at page →
Page 7
TABLEOFCONTENTS 1.0DescriptionofService 1 1 .1General 1 1.2TypesofThermometersAcceptedforCalibrationatNBS... 1 1.3ProceduresForSubmittingAThermometer forCalibration . . 2 1.4ShippingInstructions 2 2.0DesignPhilosophyandT...
Open at page →
Page 8
5.3.5BeckmannThermometers 33 5.3.6ClinicalStandardThermometers 35 5.3.7KinematicViscosityThermometers 35 5.3.8Low-temperatureThermometers 35 5.4Stem-temperatureMeasurement 36 5.5CheckingtheStandardPlatinumResistanceTherm...
Open at page →
Page 9
10.0GlassChangesandRecalibration ofThermometers 115 10.1GlassChanges 115 10.1.1TemporaryChanges 116 10.1.2PermanentChanges 117 10.2Recalibration ofLiquid-in-GlassThermometers 118 11.0Future 119 12.0Bibliography 119 vii
Open at page →
Page 11
1.0DescriptionofService 1.1General Thisserviceprovidescalibrations foravarietyofthermometers covering therangefrom-196to+538°C. Themajorityofthethermometers are liquid-in-glass butothertypesarecalibrated aswell;thermisto...
Open at page →
Page 12
Anythermometer, suchasahouseholdthermometer, consistingofapaper, plasticormetalmountingcontainingthescalegraduationsandattachedto theunmarked glassthermometer bymetalclips, isnotacceptable for calibration. Maximumself-re...
Open at page →
Page 13
NationalBureauofStandards Route270andQuinceOrchardRoad Gaithersburg, Maryland 20899 Attention: JacquelynWise Thermometerswillbereturnedviathemethodrequestedbythecustomer. If nopreference isstated,NBSwillselectabestmethod...
Open at page →
Page 14
Thebulbofthethermometer isathinglass-walled reservoir forthe thermometer liquid. Itcontainsavolumeequivalent toaspecificnumber ofdegreesofthescaledependingupontheexpansioncoefficients ofthe thermometric liquidandthebulbg...
Open at page →
Page 15
TOTALIMMERSION PARTIALIMMERSION COMPLETEIMMERSION Fig.2.Threewaysaliquid-in-glassthermometercanbeused. Apartial-immersion thermometer isdesigned toindicate temperatures correctlywhenthebulbandaspecifiedportionofthestemar...
Open at page →
Page 16
Athermometer isdesignedtobeusedatonlyoneoftheaboveimmersion depthsanditisnormallycalibratedatthatdepth. Ifthethermometer is usedatimmersiondepthsotherthanthatatwhichitwascalibratedand intendedtobeused,thetemperaturereadi...
Open at page →
Page 17
TABLE 1 TEMPERATUREEXPOSURELIMITSFORVARIOUSTHERMOMETERGLASSES [1] Strain ExposureLimits Point Continuous Intermittent °C °C °F °C °F Corningnormal7560 500 370 700 430* 805* KimbleR6 490 360 680 420 790 Jena16III 495 365...
Open at page →
Page 18
intervals ordecimalmultiples andcanbereadincorrectly. Ifthe thermometers aregraduated in0.1or0.2degreeintervals (ordecimal multiples) ,everyfifthmarkshouldbelongerthantheintermediate ones andanumbershouldappearateveryten...
Open at page →
Page 19
isalso0.4nun.Theminimuminnocaserepresentsgooddesign,andwell- designedthermometers willhavegraduation intervalsconsiderably larger thanthespecifiedminimum. Inorderforathermometer tobeusableovertheentirerangeofthescale , t...
Open at page →
Page 20
13MM 5MM 10MM 10MM (GRADUATIONABOVEISLESSTHAN100°C) '30MM(GRADUATIONABOVEISGHEATERTHAN100°C) Fig.4.Minimumdistancesforscaleplacementonaliquid-in-glass thermometer . Anyauxiliaryscalemusthavegraduations identical tothoseo...
Open at page →
Page 21
Because ofthe difficulties thatareencountered inthepractical realization oftheKTTS ,ithasbeennecessary todefineandutilize practical temperature scales. TheInternational Temperature Scaleand definingtextwasfirstadoptedin1...
Open at page →
Page 22
Intherangeoftemperature normallycoveredbyliquid-in-glassthermo- metry,theIPTS-68isdefinedbythefollowingfixedpoints: (1)Theequilibrium between theliquidandvaporphases ofoxygen (condensation point ofoxygen atonestandard at...
Open at page →
Page 23
Fig.5.StandardplatinumresistancethermometerandCutkosky ACresistance bridgeusedforcalibratingliquid-in- glassthermometers. Theconfiguration oftheresistor isinevitably theresultofcompromise betweenconflictingrequirements....
Open at page →
Page 24
Thehermeticsealthroughthesoftglassenvelopeatthethermometerhead hasbeenmadeusingshortlengthsoftungstenwire,totheendsofwhich platinumleadwiresarewelded. Theexternalplatinum leadsaresoft solderedtocopperleadsthataremechanic...
Open at page →
Page 25
Thecalibrationbathsthatuseoilasthemediumareoperatedunderahood whichisisolatedfromtheventilationsystemofthebuildingandexhausted directlyoutside. Thethermometers submitted forcalibration andthestandardplatinum resistanceth...
Open at page →
Page 26
4.2.1 IceBath Anicebath(seeFigure6)isafixed-pointapparatususedtodeterminethe referencepointonliquid-in-glassthermometers at0°C. Nostandard thermometer isneededtorealize thetemperature oftheiceandwater mixture. Ifitismade...
Open at page →
Page 27
electrical orwaterproofadhesivetapeandenclosedinametalsleeveto protecttheoperator intheeventthatitshouldbreak. Theflaskis supportedbyastructurethatwillnotallowittotipovereasily. Aclip isplacedabovetheflasktoholdthethermo...
Open at page →
Page 28
Fig.7.Steampointbath. suspended within this space. Watercondenses anddrips fromthe thermometer. Aprovisionismadebothforrelievinganyexcesspressurein thespacesurrounding thethermometers andfordetermining theexcess pressure...
Open at page →
Page 29
00 EAM i 1 Hw COo w EnH O S3 H OS S3 !=»M H W $o H Pm a wo H CO S3w OH M< H <dM COQ S3 w i— l Q S3 O * U — s U CO M 1— 1 Pi OS HH WH a CO o ;< w H 00 ON ft 1 MH \J 0) O . i j rrt J-l Q) L> H CD rrt rj-< r 4 »-i 1 rrt rrt...
Open at page →
Page 30
Fig.8.Schematicdrawingofthesteampointbath. Withabarometeraccurateto0.1mmmercury,thisprocedure iscapableof anaccuracyof0.002to0.003°C.TheFortintypebarometerwillusually sufficeforallbutthemostexactingmeasurements. 4.2.3Wat...
Open at page →
Page 31
Fig.9.Oilbath. isforcedpasttheheatingcoils,throughthebottomopeningintothe largerwell,around thethermometers, andbackintothesmallerwell throughtheconnectingpassageatthetop. Forcalibration intherangeof+1to+95°C,distilledwa...
Open at page →
Page 32
PULLEY BEARING INSULATION-^ HEATING COILS AIRSPACE SPACE BETWEEN TUBES BEARING THERMOMETER HOLDER INSULATION SUPPORT Fig.10.Drawingofwaterandoilbaths TwooilbathsaremaintainedatNBS .One,calledthelowtemperature oil bath,is...
Open at page →
Page 33
theoillevelisfirstraisedtowithin25mmofthetopofthebathby handpumpingoilfromanoverflowcollection panunderthebath. As thermometers arecalibratedatthehighertemperatures, thebathoilwill expandandleavethroughtheoverflowtubeand...
Open at page →
Page 34
V Fig.11 .Tinbath. 4.2.5Cryostat Calibrations from -1to-110 °Caremadeinthecryostat shownas Figure13.AverticalsectiondrawingisshownasFigure14.Thecryostat consistsofaninnerDewarflask,D,whichcontainsthebathliquid. This flas...
Open at page →
Page 35
0 1020 STIRDRIVE Fig.12.Drawingoftinbath. heater coils, J,woundontheoutside ofthestirrer tube. The thermometers areimmersedinsidethestirrertube,thusshieldingthemfrom theheatercoils . Thebathliquidisafive-component mixtur...
Open at page →
Page 36
Fig.13.Cryostat. usedasthebathmediumis2-methylpentane .Measuredgradients inthe cryostatdidnotexceed0.01°C.Theunknowntemperatureregionis25mm deep . 4.2.6LiquidNitrogenandLiquidOxygenPoints Acomparisonmeasurement canbemade...
Open at page →
Page 38
Fig.15.Apparatususedtocalibratethermometers attheboiling pointofnitrogenandoxygen. 5.0OperationalProcedures 5.1ExaminationandPreparationofThermometers forCalibration Everythermometer submittedforcalibration isexaminedund...
Open at page →
Page 39
dryiceisapproximately -78°C,andmercuryfreezesatapproximately -40°C,themercurywillsolidify. Thelowerhalfofthebulbiscooled andneverthestemormercurycolumn. Moderatetappingofthebulbona rubberpadorstopperorsimilarsoftspongyob...
Open at page →
Page 40
Thethirdreasonforexaminingeachthermometerunderamicroscope isto Lookforforeignmaterialinthecapillaryanddefectsinthethermometer. Theforeignmaterials mostoftenfoundareglasschipsoroxidesof mercury. Theseobjectscanholdgasorme...
Open at page →
Page 41
Ifathermometer issubmitted toNBSforcalibrationandthecalibration pointsarenotspecifiedonthepurchaseorder, itwillbetestedata referencepointandatintervalsofapproximately every100scaledivi- sions. IfitisanASTMthermometer, it...
Open at page →
Page 42
Thesethermometers arecalibrated starting atthelowesttemperature requested andproceeding insequence tothehighest point. Ateach calibration point thethermometer mustbeimmersed deeper intothe calibrationbathtomaintainthepro...
Open at page →
Page 43
5.3.4CalorimetricThermometers Insomecasesitisonlynecessary toaccuratelymeasure atemperature differenceratherthantheabsolutetemperature. Calorimetric thermometers weredesignedforthispurpose.Sincetheaccuracyofthesethermome...
Open at page →
Page 44
TABLE3 SETTINGFACTORSFORBECKMANNTHERMOMETERS Foranyscalelengthupto15° Setting Factor Setting Factor 0°C 0.9935 55°C 1.0092 5 0.9951 60 1.0102 10 0.9967 65 1.0112 15 0.9984 70 1.0121 20 1.0000 75 1.0129 25 1.0014 80 1.013...
Open at page →
Page 45
5.3.6ClinicalStandardThermometers Clinicalstandardthermometers areusedbythemanufacturers offever thermometers, whichareusedformeasuringbodytemperatures. Inthe UnitedStatestheyaremarkedindegreesFahrenheit witharangeof app...
Open at page →
Page 46
minimizestheamountofliquidleftonthecapillarywall. Sufficienttime shouldalwaysbeallowed toassurecomplete drainage. Underadverse conditionsthismaytakeanhourormore. Inaddition togooddrainage characteristics, asatisfactory l...
Open at page →
Page 47
oftheareasurrounding thebulb,whichisaisothemeantemperaturevalue oftheadjacent portion ofthethermometer stem. Stem-temperature measurements takenatNBSarebasedupontheuseoffadenthermometers wheneverpossible(seefigure19(b))....
Open at page →
Page 48
pieceoffoamisplacedatthebottomofthereentrantwell,enablingthe standardplatinumresistancethermometer tohavesomethingsoftonwhichto rest. Analuminumbushing,whichisapproximately 50mmlong,isplaced insideofthereentrantwellonthe...
Open at page →
Page 49
bulb.) Thethermometer isgentlyplacedthroughtheholderandintothe regionofloosenedice. Ifthethermometer touchesafirmsurfacebefore itisimmersedtotheimmersionlineor0°Cmark,thethermometershould beremovedandtheiceloosenedfurthe...
Open at page →
Page 50
15minutes,becausetheorganicfluidtendstoclingtothewallofthe capillary.) Whenstability isreached(themeniscusstopsmoving), the operatorgentlytapsthethermometer tofreeapossiblestickingmercury meniscusandrecordstheice-pointre...
Open at page →
Page 51
5.8PlacementofFadenandAuxiliary Thermometers forMeasurementofStem Temperatures Fadenorauxiliarythermometers, ortheircombinations, areusedtomeasure themeanemergent-stem temperature fromtheimmersionlinetothetopof themercur...
Open at page →
Page 52
n A I 209mm 'A I 78D*gr««s 15Dagraas i Fig.19(a).Measuringtheemergent-stemtemperatureofanASTM 33Cthermometerusingtwoauxiliarythermometers. 42
Open at page →
Page 53
I Fig.19(b).Measuringtheemergent-stem temperatureofanASTM 33Cthermometerusingtwofadenthermometers. 43
Open at page →
Page 54
At /K 20Degree* A I A 15Degrees Fig.20(a) .Measuringtheemergent-stem temperature ofanASTM 34Cthermometerusingoneauxiliarythermometer. 44
Open at page →
Page 55
Fig.20(b). Measuringtheemergent-stem temperatureofanASTM 34Cthermometerusingonefadenthermometer. 45
Open at page →
Page 56
cryostatandwaterbath,thetopofthebulborthetopofthecontraction chambermustbeimmersedaminimumof25mm.Intheoilbathsthebottom ofthebulbmustbeimmersed aminimumof76mmandthetopofthe contraction chamberimmersed aminimumof50mm. Int...
Open at page →
Page 57
Fig.21(a).Measuringtheemergent—stemtemperatureofanASTM11C partial-immersion thermometerwiththeimmersionline belowthesurfaceofthebathmedium. 47
Open at page →
Page 58
1\ 156Degrees X 28.2Degrees si/ I RoomTemperature 0 25mm I 50mm Fig.21(b). Placementofauxiliarythermometersneededtoobtain dataforatoo-deepimmersioncorrection. 48
Open at page →
Page 59
sametemperature asthebulb. Because ofthe50mmregionofunknown temperature nearthetopoftheoilbathsanda76mmregionofunknown temperature nearthetopofthetinbath, afadenthermometer mustbe placedinthebathtodeterminethetemperature...
Open at page →
Page 60
hi -7K 50mm 100mm V Fig.22.Measuringthetemperatureoftheportion ofatotal-immersion thermometerstemthat isintheunknown temperatureregionof theoilbathusinga100mmfadenthermometer. 50
Open at page →
Page 61
currentof1maisusedforthestandardplatinumresistancethermometer andtheamplification isadjustedsothatthevalueintheseventhdial (0.00001 fl)canbeseen. ThephaseisadjustedtozeroontheACbridge meter. Thephase/signal selectorswitc...
Open at page →
Page 62
TABLE4 SAMPLEDATASHEET ROUTINEBATCHCALIBRATION READINGS NBSTESTNUMBER CK.STD. BRIDGESPRT TO Tl T2 T3 5 445 25.46480 -0.10 -0.10 +0.03 -0.11 25.46480 19.92 19.94 19.93 19.71 27.48504 19.93 19.95 19.93 19.71 510 515 39.90...
Open at page →
Page 63
TABLE5 SAMPLEDATASHEET STEM-TEMPERATURE CORRECTIONS READINGS NBSTESTNUMBER T1 T5 T6 T7 ASTM33CASTM31CASTM11C 50MM 50MM 25MM IMMERSIONIMMERSIONIMMERSION CK.STDS. T8 T9 BRIDGESPRT 5 115 FADEN AUXILIARY 0.13 +0.5 +0.132 0.1...
Open at page →
Page 64
TABLE6 SAMPLEDATASHEET ROUTINEBATCHCALIBRATIONCOMPUTATIONS NBSTESTNUMBER CK.STD. BRIDGE SPRT TO Tl T2 T3 5 445 25.46480 -0.10 -0.10 +0.03 -0.11 25.46480 R(TP)=25.46480 a +0.10 +0.10 -0.03 +0.11 f R(0)=25.46381 b 19.92 19...
Open at page →
Page 65
wasderived frominformation given inNBSMonograph 126,Platinum ResistanceThermometry [13]andthearticle inMetrologia, TheInterna- tionalPracticalTemperatureScaleof1968AmendedEditionof1975[5]. Sincethederivationofthisfactori...
Open at page →
Page 66
A= A= A= a(l+6/100) (0.003926378)*[1+(1.496448/100)] 0.0039851342 KnowingthevalueforAandthevaluet'atthetriplepointofwater,the factorusedtoconverttheR(TP)valuetotheR(0)valuecanbedetermined asshownbelow: R(TP)/R(0)=1+At' R...
Open at page →
Page 67
TABLE7 ONEPAGEOFASTANDARDPLATINUM RESISTANCETHERMOMETERTABLE October17,1978,IPTS-68Table Temp. Resi^fcmrp J.11VC1 °C Ratio Diff . o 1000000-L•WWW 25078 1 1003986 250^86 2 1007971 25094 3 l!011955 251 !01 4 l'015938 251!0...
Open at page →
Page 68
R(t)/R(0)=1.079379 R(t)/R(0)at19°C=1.075531 Difference=0.003848 DecimalValue=Difference*Inverse Difference Decimalvalue=(0 .003848)*(252 .33) Decimalvalue=0.97 Thecalculatedtemperatureofthecalibrationbathatthefirstcalibr...
Open at page →
Page 69
.20 I Uncertainty .10 c o o a> i— l_ o O .00 1Scale Division 0 20 40 60 80 100 Temperature°C Fig.23.Thermometercorrectionversustemperature. forliquid-in-glass thermometers. Ifthesedataweretakenbytwo observers, theprobabi...
Open at page →
Page 70
n=numberofscaledegreesequivalent tothelengthofemergent stem(includingtheevaluatedregionabovetheimmersionline) t obs =observedmeantemperatureoftheemergentstem(Fadenor auxiliarythermometerreading) t sp =specifiedmeantemper...
Open at page →
Page 71
TABLE8 VALUESOFTHEDIFFERENTIALCOEFFICIENTOFCUBICALEXPANSION OFMERCURYINGLASS(k)FORMERCURY-IN-GLASSTHERMOMETERS Table8(a) MeanTemperature k k (ti+t)/2 "Normal"Glass "Borosilicate" Glass ForCelsiusThermometers 0° 0.000158...
Open at page →
Page 72
TABLE9 NBSTESTNUMBER T4 T5 T6 T7 ASTM33CASTM34CASTM11C 50ran 50ran 25run IMMERSIONIMMERSIONIMMERSION -K3.U3 +0.5 - .5 .0 -0.43 SAMPLEDATASHEET STEM-TEMPERATURE CORRECTIONS COMPUTATIONS CK. T8 STDS. T9 -0.5 +0.132 +0.1* -...
Open at page →
Page 73
100.52 100.40 t=100.15°C35,47688 100.52 100.10 698 707 -.37 -.25 R(t)=35.17698 R(t)/R(0)=1.393231 150.1 148.82 t=148.39°C40.19077 103rrm 100nm 150.1 148.82 096 112.9 140.0 -1.7 -.43 106 +.6 0.0 +1.7 e -.03 R(t)=40.19093...
Open at page →
Page 74
Thefourthvalueisthespecifiedaveragetemperatureoftheemergentstem (t sp ).Thisvalueisthetemperaturethattheemergentstemshouldbe. If thethermometer isatotal-immersionthermometerhavingaportionofthe stemoutofthemediumbeingmeas...
Open at page →
Page 75
TABLE10 FADENTHERMOMETERCORRECTIONS FadenThermometerLength 100mm 103mm 114mm 200mm Ice-Point Correction 0.0 +0.6 +0.9 +1.0 ScaleCorrection Temperature °C 0 0,0 0.0 0.0 0.0 50 +0,.1 -1.0 -2.1 -0.3 100 +1..2 -1.0 -2.4 +0.2...
Open at page →
Page 76
Calculating thestem-temperature correctionafterthetemperature ofthe emergentstemisdetermined isidenticalwiththeonediscussedinSection 6.3.1. Thecalculation isgivenbelow: EmergentStem-temperaturecorrection=kn(t obs -t sp )...
Open at page →
Page 77
6.3.5Too-deepImmersionCorrection Sometimes itisnecessary toimmerseapartial-immersionthermometerwith theimmersionlinebelowthesurfaceofthebathmedium. Thiswillbethe caseifthedistancefromtheimmersionlinetothebottomofthetherm...
Open at page →
Page 78
Ifthecomputedcorrection ispositive, itisaddedtothethermometer correction. Ifthecomputedcorrection isnegative, itissubtractedfrom thethermometer correction. Theresultingcorrection isapplicablewhen thisthermometer isusedat...
Open at page →
Page 79
tableisgiveninTable8(b),wherethevalueofkcorresponding tothe bathtemperatureandfadenthermometerreadingcanbeeasilydetermined. ThevalueLBisequivalent ton,asdiscussed inSections6.3.1thru 6.3.4. Thisvariationoftheformula isus...
Open at page →
Page 80
6.3.8EmergentStem-temperatureCorrectionforBeckmannThermometers Beckmannthermometers arealsocalibrated attotalimmersion, butmaybe usedatpartial immersion. Theformulausedtocompute theemergent stem-temperature correction is...
Open at page →
Page 81
U.S.DEPORTMENTOFCOMMERCE NOTIONALBUREAUOFSTANDARDS NATIONALMEASUREMENTLABORATORY GAITHERSBURG,' MD.£0899 REPORTOFCALIBRATION LIQUID-IN-GLASS THERMOMETER. TESTEDFOR:NATIONALBUREAUOFSTANDARDS DIVISION522 MARKED:SURETYTO RA...
Open at page →
Page 82
PAGE£ THETABULATEDCORRECTIONSAPPLYPROVIDEDTHEICE-POINTREADING,TAKEN AFTEREXPOSUREFORNOTFEWERTHAN3DAYSTOATEMPERATUREOFABOUT23 DEGREESC(73DEGREESF),IS-0.10C .IFTHEICE-POINTREADINGISFOUND TOBEHIGHER(ORLOWER)THANSTATED,ALLOT...
Open at page →
Page 83
1)Company,agency,orestablishment thatsubmitted thethermometer, 2)Manufacturer's tradenameandserialnumberofthe thermometer , 3)Rangeandgraduationinterval, 4)Calibrationtemperaturesandcorrections, 5)Temperaturescaleused(IP...
Open at page →
Page 84
CORRECTIONFOREMERGENTSTEM Thisstemcorrectionsheetisdesigned toexplain theprocessofcomputingthecorrectionsforemergent stem,buthasnotbeenspecificallyadaptedtothe individualthermometerwithwhichitisissued. Ifatotal-immersion...
Open at page →
Page 85
DIFFERENTIALCORRECTIONFOREMERGENTSTEM Thissheetisdesignedtoexplainaconvenientprocess ofcomputingthedifferentialcorrectionforemergent stembutliasnotbeenspecificallyadapted tothe individualthermometerwithwhichitisissued. I...
Open at page →
Page 86
heatedtoahighertemperature, anerroror0.01°Corless,foreach10 degreedifferencebetweenthetwotemperatures, maybeintroduced. The tabulatedcorrections applyifthethermometer isusedinitsupright position; ifusedinahorizontalposit...
Open at page →
Page 87
Thissheetgivesatableofsettingfactorsandanexampletoexplaintheprocesses employed incalculations.Theexample,however,hasnotbeenspecificallyadaptedtothe individualthermometerwithwhichthissheetisissued. Thethermometer issaidto...
Open at page →
Page 88
Othernotesthatareoccasionallyusedandareselfexplanatoryaregiven below: Thisthermometer complieswiththeaccuracyrequirement oftheapplicable ASTMspecification totheextentofthetestsmade. Thisthermometer doesnotcomplywiththeac...
Open at page →
Page 89
thefirstandthirdplatinumresistancethermometervaluesforthefirst liquid-in-glass thermometer. Checkanyfadenorauxiliary thermometer readingsforcorrectness. 8.1UseofCheckStandards Theabovesourcesoferrorcanbepracticallyelimin...
Open at page →
Page 90
TABLE11 RECORDOFCHECKSTANDARDCORRECTIONS ThermometerSerialNumber 274762 DATE 3/26/76 3/29/76 4/1/76 4/6/76 4/21/76 5/17/76 5/25/76 BOOK F115 C193 C193 C193 G30 C194 F115 PAGE 182-83 169 173 174-75 192-93 2 188 Temp. Corr...
Open at page →
Page 91
onthecheckstandardtobeabletotellwhetheritisthenewortheold valuethatisout-ofcontrol. Lackofagreement isalwaystakenasan indicationofaproblemwiththecurrentcalibration, andcalibrations on allthermometers inthegrouparerepeate...
Open at page →
Page 92
TABLE12 TOLERANCESANDUNCERTAINTIES FOR LOW-TEMPERATURETOTAL-IMMERSION THERMOMETERS Temperature Typeof Graduation Toler- Uncertainty rangein thermometer intervalin ancein indegrees degrees degrees degrees Celsiusthermomet...
Open at page →
Page 93
TABLE13 TOLERANCESANDUNCERTAINTIES FOR LOW-TEMPERATUREPARTIAL-IMMERSION THERMOMETERS Temperature rangein degrees Typeof thermometer Graduation intervalin degrees Toler- ancein degrees Uncertainty indegrees Celsiusthermom...
Open at page →
Page 94
TABLE14 TOLERANCESANDUNCERTAINTIES FOR CELSIUSTOTAL-IMMERSION MERCURYTHERMOMETERS Temperaturerangein degrees Graduation intervalin degrees Tolerance in degrees Uncertainty indegrees Thermometersgraduatedunder150°C 0upto1...
Open at page →
Page 95
TABLE15 TOLERANCESANDUNCERTAINTIES FOR FAHRENHEITTOTAL-IMMERSION MERCURYTHERMOMETERS Temperaturerangein degrees Graduation intervalin degrees Tolerance in degrees Uncertainty indegrees Thermometersgraduatedunder300°F 32u...
Open at page →
Page 96
TABLE16 TOLERANCESANDUNCERTAINTIES FOR CELSIUSPARTIAL-IMMERSION MERCURYTHERMOMETERS Temperaturerangein degrees Graduation intervalin degrees 3 Tolerance in degrees Uncertainty indegrees 13 Thermometersgraduatedunder150°C...
Open at page →
Page 97
TABLE17 TOLERANCESANDUNCERTAINTIES FOR FAHRENHEITPARTIAL-IMMERSION MERCURYTHERMOMETERS Temperaturerangein degrees Graduation intervalin degrees 3 Tolerance in degrees Uncertainty indegrees 13 Thermometersgraduatedunder30...
Open at page →
Page 98
reason. Therefore, increasing thelengthofadegreeonthescale,for practicalbulbsizes,improvesthermoraetric performance onlytoacertain point. Beyondthispointprecisionofreadingmayreadilybemistakenfor accuracyintemperature mea...
Open at page →
Page 99
Overthepastsixteen years, fourteen liquid-in-glass thermometers, representing themostaccuratetypesinagiventemperature range,have beenusedascheckstandards. Twototal-immersionthermometers ofeach typewerechosenforthestandar...
Open at page →
Page 100
madeforthebulbvolumechange. Allofthedatawerenormalized toa constant ice-point readingrepresenting anapproximate averageofthe ice-pointcorrections. Thedataforselectedpointswereagainplottedand theresultsaregiveninFigures29...
Open at page →
Page 101
0.00 YEAR Fig.29(a) Plot forthermometer, serial number T531432, atfive temperatures (1at32°F,2at-30°F,3at-40°F,4at -50°F,and5at-60°F) .Thereadingswerenotcorrectedfor ice-pointshifts. 0.00 -0.05- -0.10 -0.IS - -0.20 - -0....
Open at page →
Page 102
0.00 72 73 74 75 76 77 78 79 80 81 YEAR Fig.30(a). Plot for thermometer, serial number T531438, attive temperatures (1at32°F, 2at-30°F,3at-40°F,4at -50°F,and5at-60°F) .Thereadingswerenotcorrectedfor ice-pointshifts. 72 7...
Open at page →
Page 103
0.18 0.16 0.14 S0.12 0.10 0.08 0.06 YEAR Fig.31(a). Plot for thermometer, serial number 274762, at six temperatures (1at0°C,2at10°C,3at20°C,4at30°C,5 at40°C,and6at50°C) .Thereadingswerenotcorrected forice-pointshifts. 0....
Open at page →
Page 104
0.20 0.18 0.16 0.14 0.12- 0.10 72 74 76 78 80 82 YEAR 32(a) u LU a. <Y a u cc UJ I- Lu E o a. LU I H Plot for thermometer, serial number 275764, at six temperatures (1at0°C,2at10°C,3at20°C,4at30°C,5 at40°C,and6at50°C) .T...
Open at page →
Page 105
0.04 -0.04 | I I I I | I I I I | I I I I | I I I I | I I I I 72 73 74 75 76 77 YEAR Fig.33(al).Plot for thermometer, serial number 4030424, atfour temperatures (1at0°C,2at50°C, 3at60 ° ,and4at 70°C) .Thereadingswerenotco...
Open at page →
Page 106
u o .100 .075 .050 .025 .000 7025 7G 77 33(b) Plot forthermometer, serial number 4030424, atthree temperatures (1at0°C,2at70°C,and3at100°C) .The readingswerecorrectedforice-pointshifts. 0.01 0.00 -0.01 o |-0.0E -0.03 34(...
Open at page →
Page 107
0.08 0.06 0.04 0.02 0.00 -0.03 Fig.34(a2) YEAR Plot forthermometer, serial number 4030425, atthree temperatures (1at80°C,2at90°C,and3at100°C) .The readingswerenotcorrectedforice-pointshifts. .075 .050 .025 ,000 -025- -05...
Open at page →
Page 108
ijJ 0.2 0.1 0.0 0.1 -0.2 35(al) u LU or <r a u a. LU h- lll £ o £ LU X Plotforthermometer, serialnumber48425,atsixtemperatures (1at0°C,2at10°C,3at20°C,4at30°C,5at40°C, and6at50°C) .Thereadingswerenotcorrectedforice- poin...
Open at page →
Page 109
0.le -0. 15— J 1 1 1 1 1 1 1 76 78 80 82 84 35(a2).Plot for thermometer, serial number 48425, at five temperatures (1at60°C,2at70°C,3at80°C,4at90°C, and5at100°C) .Thereadingswerenotcorrectedforice- pointshifts. 0.05-1 1...
Open at page →
Page 110
0.30 77 78 79 80 81 82 83 84 VEPR 36(al).Plot for thermometer, serial number 198694, at six temperatures (1at0°C,2at10°C,3at20°C,4at30°C,5 at40°C,and6at50°C) .Thereadingswerenotcorrected forice-pointshifts. 77 78 79 80 8...
Open at page →
Page 111
0.12 0.10 0.08 YEAR 36(b). Plot for thermometer, serial number 198694, atthree temperatures (1at0°C,2at40°C,and3at90°C) .The readingswerecorrectedforice-pointshifts. -0.1 -0.5 37(al).Plot for thermometer, serial number,...
Open at page →
Page 112
o -0.35 -0.40 -0.45 -0.50 -0.55 -0.60 37(a2) Plot for thermometer, serial number 382213, at six temperatures (1at150°C,2at160°C,3at170°C,4at 180°C,5at190°C,and6at200°C) .Thereadingswerenot correctedforice-pointshifts. -0...
Open at page →
Page 113
0.0 72 74 76 78 80 82 84 YEAR 38(al).Plot for thermometer, serial number 382215, at six temperatures (1at0°C,2at100°C, 3at110°C,4at 120°C,5at130°C,and6at140°C) .Thereadingswerenot correctedforice-pointshifts. -0.2 72 74...
Open at page →
Page 114
-0. 1 -0.2 -0.3 -0.4 -0.5 Fig.38(bl).Plot for thermometer, serial number 382215, atthree temperatures (1at0°C,2at100°C,and3at200°C) .The readingswerecorrectedforice-pointshifts. -0.20 -0.E5 -0.30 -0.35 -0.40 -0.45 -0.50...
Open at page →
Page 115
0.1 0.0 -0.1 -0.2 -0.3 Fig.39(al) Fig.39(a2) Plot forthermometer, serial number T1124411, atfive temperatures (1at0°C,2at200°C, 3at210°C,4at 220°C,and5at230°C) .Thereadingswerenotcorrectedfor ice-pointshifts. 0.30 0.25 0...
Open at page →
Page 116
0.25 0.20 0.15 0.10 0.05 YEAR Fig.39(b). Plot for thermometer, serial number T1124411, attwo temperatures (1at0°Cand2at250°C) .Thereadingswere correctedforice-pointshifts. or tr a u <r HI h- £ a E ir uj r 0.2 0.1 0.0 -0....
Open at page →
Page 117
0.35--- 0.15 Fig.40(a2).Plot for thermometer, serial number T1124412, atsix temperatures (1at250°C,2at260°C,3at270°C,4at 280°C,5at290°C,and6at300°C) .Thereadingswerenot correctedforice-pointshifts. 0.25 0.20 0.15 W 0.10...
Open at page →
Page 118
1.0 -1.5 41(al).Plot for thermometer, serial number 234453, atfour temperatures (1at0°C,2at300°C,3at310°C,and4at 320°C) . Thereadings werenotcorrected forice-point shifts . -0.75- -1.00 -1.25 g-1.50 1.75 -2.00 YEAR 41(a2...
Open at page →
Page 119
0.5 41(b). Plot for thermometer, serial number temperatures (1at0°Cand2at350°C) . correctedforice-pointshifts. 234453, at two Thereadingswere 42(al).Plot for thermometer, temperatures (1at0°C 330°C,and5at340°C) . ice-poi...
Open at page →
Page 120
-a. s 74 76 78 80 83 TEAR Fig.42(a2).Plot for thermometer, serial number 234456, atfive temperatures (1at350°C,2at360°C,3at370°C,4at 390°C,and5at400°C) .Thereadingswerenotcorrectedfor ice-pointshifts. Acomparisonofthecal...
Open at page →
Page 121
TABLE20 LIMITSTORANDOMERRORCOMPUTEDFROMCHECKSTANDARDMEASUREMENTS Temperature Range (°F) Bath Temperature (°F) Check Standard Number Limitto RandomError (3Standard Deviation) Degreesof Freedom 60to-30 -30 -40 -50 -60 -30...
Open at page →
Page 122
Temperature Range (°C) 0to100 Bath Temperature Check Standard Limitto RandomError Degreesof Freedom 100to200 ( °C) Number (3Standard Deviation) 10 48425 .030 22 20 48425 .031 44 30 48425 .036 38 40 48425 .025 42 50 48425...
Open at page →
Page 123
Temperature Range CO 200to300 Bath Temperature CO 200 210 230 250 270 300 210 240 250 260 270 290 300 Check Standard Number T112411 T112411 T112411 T112411 T112411 T112411 T112412 T112412 T112412 T112412 T112412 T112412...
Open at page →
Page 124
TABLE22 CALCULATEDUNCERTAINTIES FORLIQUID-IN-GLASSTHERMOMETERS Temperature Range (°F) -60to-30 Thermometer Graduation (°F) 0.2 Limitto RandomError (3a) 0.054 Total Systematic Error 0.029* Calculated Uncertainty 0.083 Tem...
Open at page →
Page 125
fluidclingstothecapillarywallastheliquidrecedesintothebulb. Becauseofthisthethermometer shouldnotbereadimmediately after insertion intothemedium. Uptoonehourmightberequiredbeforea readingcanberecorded. Datawillhavetobeta...
Open at page →
Page 126
closeapproachtoquasiequilibriummaybereachedintheorderofhoursat annealingtemperatures, whileinfinitetimemayberequiredatmuchlower temperatures . Ifaglassthathasbeenheatedtoahightemperature isallowedtocool rapidly, itwillbe...
Open at page →
Page 127
measureagiventemperature, itwillindicateanerroneously lowvalueif ithas,withinashortperiodoftime,previouslybeenexposedtoahigher temperature. Withthebettergradesofthermometric glassestheerror resultingfromthishysteresiswil...
Open at page →
Page 128
10.2Recalibration ofLiquid-in-GlassThermometers Thethermometer scalecorrectionsquotedintheNBSReportofCalibration applyaslongastheice-pointreadingremainsthesameasobservedduring thecalibration. Asdiscussedabove,subsequentc...
Open at page →
Page 129
currentsetofcorrections forthethermometer (seeNBSMonograph150[7], Section4).Aliquid-in-glassthermometerwillneverhavetobereturned toacalibrationlaboratory, suchasNBS,forrecalibration . 11.0Future Thedemandforliquid-in-gla...
Open at page →
Page 130
Ween,Sidney,"TheBeckmannDifferentialThermometer: ItsPrinciples ofConstruction andApplication," MaterialsResearchandStandards, MTRSA,Vol.12,No.8,August1972,p.31. Buckingham, E. ,Thecorrection foremergentstemofthemercurial...
Open at page →
Page 131
NBS-114A(REV.2-8C) U.S.DEPT.OFCOMM. BIBLIOGRAPHICDATA SHEET(Seeinstructions) 1.PUBLICATIONOR REPORTNO. NIST/SP-250/23 2.PerformingOrgan.ReportNo 3.PublicationDate September1938 4.TITLEANDSUBTITLE Liquid-In-Glass Thermome...
Open at page →
Page 135
THESP250SERIESONNISTMEASUREMENTSERVICES* SP250-1 SpectralRadianceCalibrations SN003-O03-02792-8 $3.50 SP250-2 FarUltravioletDetectorStandards PB87227609 SP250-3 RadiometricStandardsintheVacuum Ultraviolet SN003-003-02806...
Open at page →
Page 136
U.S.DepartmentofCommerce NationalInstituteofStandardsandTechnology (formerlyNationalBureauofStandards) Gaithersburg,MD20899 OfficialBusiness PenaltyforPrivateUse$300 MlOP StimulatingAmerica'sProgress
Open at page →