High Vacuum Standard and Its Use
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
NISTMeasurementServices: NEWNISTPUBLICATION April20,1989 HighVacuumStandard andItsUse SharrillDittmann U.S.DepartmentofCommerce NationalInstituteofStandardsandTechnology (formerlyNationalBureauofStandards)
Open at page →
Page 2
CenterforBasicStandards TheCenterforBasicStandardsdevelopsandmaintainsthescientificcompetenciesandlaboratoryfacilitiesnec- essarytopreserveandcontinuetorefinethebasephysicalquantitiesuponwhichtheNation'smeasurement syste...
Open at page →
Page 3
NISTMEASUREMENTSERVICES: HighVacuumStandardandItsUse SharrillDittmann CenterforBasicStandards NationalMeasurementLaboratory NationalInstituteofStandardsandTechnology Gaithersburg,MD20899 March1989 U.S.DEPARTMENTOFCOMMERC...
Open at page →
Page 4
LibraryofCongressCatalogCardNumber.88-600605 NationalInstituteofStandardsandTechnologySpecialPublication250-<34 Natl.Inst.Stand.Technol.Spec.Publ.250-34,68pages(March1989) CODEN:NSPUE2 U.S.GOVERNMENTPRINTINGOFFICE WASHIN...
Open at page →
Page 5
PREFACE CalibrationsandrelatedmeasurementservicesoftheNationalInstituteof StandardsandTechnologyprovidethemeansformakersandusersofmeasuring toolstoachievelevelsofmeasurementaccuracythatarenecessarytoattain quality,produc...
Open at page →
Page 6
documentsmaybedirectedtotheauthorortooneofthetechnicalcontact personsidentifiedintheUsersGuide(SP -250) . Suggestionsforimprovingtheeffectiveness andusefulnessofthenewseries wouldbeverymuchappreciatedatNIST. Likewise,sug...
Open at page →
Page 7
TABLEOFCONTENTS Page Preface iii I.Introduction 1 II.DescriptionoftheNISIVacuumCalibrationService 1 III.Theory,DesignPhilosophyandDescriptionoftheCalibration System 2 TheChamber 3 TheFlowmeter 4 TheOrifice 6 ExtensiontoL...
Open at page →
Page 8
TABLEOFCONTENTSrCONTINUED) Appendices A.SampleReportofCalibrationforIonizationGage B.SampleReportofCalibrationforMolecularDragGage G.DimensionsandConductanceoftheOrificeintheNIST DynamicExpander vi
Open at page →
Page 9
LISTOFTABLES Page I.Stunmaryofthevacuumcalibrationservice 2 II.Primarystandardconfigurations usedforcalibrations 15 III.Primarystandardnonllnearltles 25 IV.Uncertainties oftheNISThighvacuumstandardwithnitrogen 29 V.Resul...
Open at page →
Page 11
I INTRODUCTION Vacuummeasurements areanimportanttoolformanysegmentsofthe AmericaneconomyandforAmericandefenseefforts. Theaccuracyofthese measurementsaffectsalargevarietyofindustriesrangingfromspacesimula- tiontosemicondu...
Open at page →
Page 12
TABLEI.SummaryoftheVacuumCalibrationService Sensor Calibration Range Standard Uncertainty Pa percent MDG 5x10-3 Primaryhigh 1.4-1.6 vacuum IG IG 10"^ -10-1 10-6 .iQ-t, CalibratedMDG Primaryhigh vacuum 1.4-2.6 1.4-5.2 Ifc...
Open at page →
Page 13
THECHAMBER Thedesignofthevacuumchamber,illustrated infigure2,islargely determinedbytheneedtomaintainalowbackgroundpressure, toprovideroom tomountthegagestobecalibrated, andtomaintainageometrywherethe orificeconductanceca...
Open at page →
Page 14
thelowerchamberandbythe15-cmi.d,connectionbetweenthelowerchamber andthepijmp. Theturbomolecular pumpwaschosenforitsstablehighpumping speed,lowbasepressure,andcompatibilitywithallgases. Ultra-highvacuumconstructionpractic...
Open at page →
Page 15
afterasufficiently longtimefollowingestablishment ofasteadygasflow. Second,stablepressuresarereadilygeneratedforlongperiodsoftime. Third,theonlydimensionswhichneedtobecharacterized arethoseofthe orificeandthevolumeofthep...
Open at page →
Page 16
Thenextstepistoisolatetheworkingreservoir (E)fromtheballast volumesothatflowisoutoftheworkingreservoiraloneintothechamber. Thisflowofgaswilltendtoreducethepressureintheworkingreservoirbut constantpressureismaintainedbyad...
Open at page →
Page 17
uppersidetoadepthof0.61cmfromthecenterouttoadiameterof8.89cm, andacentered1.12-cmdiameterholewasboredthroughtheplate. The undersideoftheplatewasmachinedtoformahollow,truncated80°half-angle cone,leavingathicknessof0.74mma...
Open at page →
Page 18
thepressureisreduced,therandomerroroftheMDG,typically10"^Paor greater,precludesitsuseforadirectdetermination ofRp .Rpwillbea functiononlyoftheorificeconductanceandthepumpspeed,andinthe molecularflowrange,theonlyvariablec...
Open at page →
Page 19
lessthan1%althoughthismaybefortuitoussincefluxandpressuregradients ofseveralpercentareknowntoexistinthelowerchamber. Theflowratiocannotbedirectlymeasuredforpressureslowerthanthose thatcanbereliablygeneratedbyameasuredflo...
Open at page →
Page 20
MDGperformanceatpressuresabove10"^Paislimitedbynonlinearities arisingfrominteractionsbetweenthegassurroundingtheballandthewalls ofthesteelthimblewhichcontainstheballinthevacuumchamber. The resultingchangesmaybeempiricall...
Open at page →
Page 21
COMPONENTSELECTIONCRITERIA Eachpartofthecalibrationsystemmustmeetcertainperformance criteriatoinsurethatthesystemisoperatingunderoptimumconditions .The chamberandorificecriteriaarediscussedabove. Thethreeremainingparts o...
Open at page →
Page 22
Theperformanceofthesegagesisdiscussedinthesectiononerrors. Withcarefuluse,thecalibrationofthegagewillchangenomorethan2%over a2-yearperiod. TheFlowmeter ,Allletterreferencesinthissectionaretofigure1. TheGases Becausetheor...
Open at page →
Page 23
pistons,respectively. Thepistonsarecoupledtosteppingmotorsthrough bellowsandtwenty-turns-per-inchmicrometerscrews. Thereisabout one-quarterturnbacklashinthepistondrivebutthishasnoeffectasthe pistonisalwaysdriveninthesame...
Open at page →
Page 24
Theflowmeteristhermallyinsulatedbutnottemperature-controlled. It isinsulatedwithpolystyrene toprotectitfromshort-termtemperature fluctuationsandinsureinternaluniformity. Asecondquartzthermometer is locatedinsidethepolyst...
Open at page →
Page 25
foundturbomolecular pumpstobesatisfactory iftheMDGsaremountedwith care . IV.OPERATIONALPROCEDURESFORCALIBRATIONUSING THENISTHIGHVACUUMSTANDARD Thissectiondescribesinsomedetailtheoperationoftheorifice-flow standard. Users...
Open at page →
Page 26
INSTALLATIONANDPREPARATIONOFGAGES Bake-Out Eachventingofthesystemmustbefollowedbyacompletebakingofthe chamberandinletlines. Thechamberisdesignedtobebakedattemperatures upto250°C.Thegaspisonflowmetermaynotbebakedbecauseof...
Open at page →
Page 27
becausetheinletlineisgroundedattheflowmeterandatthechamber. The inletlineshouldsizzlewhenwaterisdroppedonit.Inaddition, theheater ontheturbomolecular pumpshouldbestarted. Theovenshouldremainattemperatureforatleast8hours....
Open at page →
Page 28
checkedandadjustedifnecessaryandthezerosoftheelectrometers shouldbe setbymechanicaladjustmentonahigh-rangescaleandelectricaladjustment atlow-rangescalereadingsafterremovingthecollectorlead. Forpressuresatthelowendofthera...
Open at page →
Page 29
Pa.Then,gasmaybeintroducedintotheappropriatehalfofthechamberfrom theflowmeter. Dependingonthecalibrations tobedone,theflowmetermaybeusedto generateaknownflowofgasoritmaybeusedsimplytoestablishasteady flow.Thesecondcaseis...
Open at page →
Page 30
finalreadingshouldbeslightlypositive. Oncethevalveisclosed,the pressureintheworkingreservoirwillstarttofallbecausegasisflowing outofthereservoirintothechamber. Anytimebeforethedifferential pressurereadingbecomesnegative,...
Open at page →
Page 31
chamberandthegagesaremonitoreduntilthereadingsaresteady. Atthis timethegagesareread:MDGreadingsarestarted,theIGsarereadonce,and thentheIGsarereadinreverseorder. Theanalysisofthegagereadings is describedbelowwiththeexcept...
Open at page →
Page 32
Thereferencereservoirtemperaturebeforeandaftertherun, . Whentheflowisintotheupperchamber,theformulausedtocalculatethe pressureis Pu =(Q/C)(Rp)/(Rp-l) (3) whereQisthethroughputintotheupperchambergivenby Q=Pf,(dV/dt)(T,h/T...
Open at page →
Page 33
S=(I, -I,„)/(I,P) (13) (14) where R,,isaresistorspecifictothecontrollerandtheemissioncurrent range, isthevoltagedropacross ,1^istheemissioncurrent, isthe measuredcollectorcurrent, EisarangefactorfortheelectrometerandPis...
Open at page →
Page 34
sincetheorificeisanidealsink. Thiscontributionwilldecreasethe pressureatthegageportsfromthatgivenineq(7)byafactorof0.99956. Thesecondcontribution ismoredifficulttocalculatebecausethe scatteringoftheinletflowproducesadiff...
Open at page →
Page 35
resultsarebasedonthereadingsof12MDGstakenoveraperiodofkdaysand aresummarizedintableIIIbelow. Totaldeviationsareexpressedasthemean differencebetweenthepressureindicatedbytheMDGsandpressuregenerated bythehighvacuumstandard...
Open at page →
Page 36
pressuresassumesthatpumpspeeddoesnotchangewithdecreasingpressure. Anychangeinthepumpspeedwillcauseerrorsinthecalculatedpressure, withcorresponding errorsinthesensitivityofanygagecalibrated. In particular, adecreaseinpump...
Open at page →
Page 37
RANDOMERRORS Upperboundsontherandomerrorscanbesetbyevaluatingthevariation ofrepeatedgagecalibrations. Thesenumberswill,ofcourse,includethe randomerrorsofthegagesaswellasthoseofthestandard. Ameasureoftheshort-termrepeatab...
Open at page →
Page 38
Amorerealisticassessmentoftherandomerrorscanbeobtainedfrom calibrationsrepeatedondifferentdays. Forflowintotheupperchamberwe expectthaterrorsatthelowestpressureswillbedominatedbyrandom flowmetererrors. Thedatainfigure4a)...
Open at page →
Page 39
ThetabulatedrandomerrorsareupperboundsbasedonrepeatedMDG calibrationswithnitrogenandargon. Theseincludedtherandomerrorsofthe standard,flowmeterandgages. Thevalueat 10"'' Paisbasedonthreetimes theshort-termdeviationsobser...
Open at page →
Page 40
VI.OOALITYCONTROL Thedeterminationofsystematicandrandomerrorsinaprimarystandard isbutthefirststepincharacterizing itsperformance. Byitsverynature, aprimarystandardisdifficulttomonitoroverlongperiodsoftimeand variouslabor...
Open at page →
Page 41
Bourdongage. Someinitialworkhasbeendonetocomparethetwoflowmeters usingbothstandardleaksandMDGsastransferstandards. Theleakworkhas beendoneatlowflowrateswhiletheMDGworkhasofnecessitybeendoneat muchhigherflowrates. ElevenM...
Open at page →
Page 42
AsecondsetoffourballswassenttoNISTin1986inacontinuation of theroundrobin. PreliminaryresultsindicatethatthemeanofthefourNIST valuesforargonwaswithin0.5%ofthepilotlaboratory'svalue. Theresults forninelaboratories (represe...
Open at page →
Page 43
IonGageComparisons IongagesarenotasstableasMDGsandwedonotexpecttogainasmuch informationaboutsystemperformancefromouriongagedata. However,because ofthebroadpressurerangecoveredbythesegages,wecanusethem tolook forsystemati...
Open at page →
Page 44
of1-2%in2year'stime.Thesepa^rticularcustomersareknowledgeable users ofthegageandweexpectthattherotorshavebeentreatedcarefullybetween calibrations. VII.NEEDSANDPLANS Themajorlimitationsofthissystemare1)thepresenceofelasto...
Open at page →
Page 45
automation,amongthosebeingtherecordingofthepistondriveinformation, therecordingofCDGandtemperatureinformation, andtherecordingofMDG transferstandardreadings. ThoseIGswhicharetestedonNIST-designed controllersmayalsoberead...
Open at page →
Page 46
Martinforhiseffortsinthemanufactureandassemblyofthemanycomponents involved. IX.REFERENCES [I] K.E.McCulloh, C.R.Tilford, C.D.Ehrlich, andF.G.Long,"Low-Range FlowmetersforUsewithVacuumandLeakStandards," J.Vac.Sci.Technol....
Open at page →
Page 47
[15]P.L.M.Heydemann, C.R.Tilford,andR.W.Hyland,"Uitrasonic ManometersforLowandMediumVacuaunderDevelopmentattheNational BureauofStandards," J.Vac.Sci.andTechnol. 14,597(1977). [16]R.W.Hyland,"NISTMeasurementServices:Calib...
Open at page →
Page 48
VacuumGageCalibration : "CalibrationofSpinningRotorGasFrictionGaugeAgainstaFundamentalVacuum PressureStandard," G.Comsa,J.K.Fremerey,andB.Lindenau(KFA)and G.MesserandP.Rohl(PTB) ,J.Vac.Sci.Technol. 17,642(1980). "TheCali...
Open at page →
Page 49
Gas Inlet Line Gas Supply Cylinder n Ballast Volume A Working Reservoir Variable Leak H 1 to Chamber u Reference Volume B Figure1.SchematicofthePrimaryHighVacuumStandardIncludingthe Chamber,theFlowmeterandtheOrifice. Let...
Open at page →
Page 50
Valve Gas Bakeable OrificePlateLifting Mechanism Baffle Orifice Plate Baffle Bakeable Valve ) Spinning Rotor Gage ^Bakeable Valve To Turbopump Figure2.SchematicoftheVacuumChamberShowingtheLifting ^ MechanismandSealingMet...
Open at page →
Page 51
Figure3.AxialSectionoftheOrificePlate,withDimensions inInches. DetailsoftheCentralDuctareShowninfigure7.ThisPlate canbeRaisedtoProvideaLargeConductanceBetweentheUpper andLowerChambersoftheDynamicExpanderDuringEvacuation....
Open at page →
Page 52
Figure4.Representative MeasuredLow-PressureNitrogenSensitivities for DifferentIonGages. Sensitivities forEachGageareRelativeto itsValueat10'*Pa.The+'sIndicateFlowintotheLower Chamber,theO'sareintotheUpperChamber. TheData...
Open at page →
Page 53
Figure5.DeviationsfromtheInitialValueoftheEffectiveAccommodation CoefficientofFiveMolecularDragGageRotorsasDeterminedby CalibrationsAgainsttheNISTPrimaryHighVacuumStandard. All DatawereTakenUsingNitrogenGas.Valueshavebee...
Open at page →
Page 54
0.10 (0 a I0.05 - c (0 -5 -4 -3 LogPressure,Pa 0.06 (0 Q. 0.05 0) I0.04 XXXX 5 X >, X X * tA? 0.03 -8 -6 -5 -4 -3 LogPressure,Pa a)andb)IonGageSensitivity inNitrogenforTwoDifferent UltraHighVacuumIonizationGages'asDeterm...
Open at page →
Page 55
I i (0,0,-a) I I z Figure7.DetailoftheCentralDuctThroughtheOrificePlate. The SurfacesMarkedAandBareSphericalSegmentsLappedto RadiusofCurvatureR,andTheirCentersofCurvatureare MarkedwithCoordinates (0,0,-a)and(0,0,a).Respe...
Open at page →
Page 57
XI.APPENDICES APPENDIXA SAMPLEREPORTOFCALIBRATIONFOR IONIZATIONGAGE IonGageTubeandController Submittedby CUSTOMERNAME CITY,STATEZIP TPDCalibrationNumber: HVXX NISTIdentification ofGage:XXX SupplierofGage:SUPPLIER GageMod...
Open at page →
Page 58
Pa.Usingcalibratedmoleculardraggages,theestimateduncertainty inthe generatedpressurebetween 10"'' and10" •'• Pais1.6%±2x10"^Pa, Thegagetubewasmountedonthecalibrationchamberandevacuatedand bakedovernightat250"C.Itwasdegas...
Open at page →
Page 59
Theoperatingparametersoftheelectrometerwereindependently checked andarenotedbelow,foryourinformation. Thecontrollerrequired approximately 24hoursafterbeingturnedontoreachastableindication. CurrentatCollector Input.Amps I...
Open at page →
Page 61
APPENDIXB SAMPLEREPORTOFCALIBRATIONFOR MOLECULARDRAGGAGE SpinningRotorVacuumGage Submittedby CUSTOMERNAME CITY,STATEZIP TPDReferenceNumber: HVX NISTIdentification ofFlange: 654321 Manufacturer's Identification ofFlange:...
Open at page →
Page 62
Atthetimeoftest,thepickupsignalfromtheball,theresidualdrag andthefrequencydependenceoftheresidualdragwereallmonitored. Allof thesefactorsmaychangeduringuseandmustberedetermined atthetimeof use.Theballwasremagnetizeduntil...
Open at page →
Page 63
Theeffectiveaccommodationcoefficientgivenfortheballappliesonly fornitrogen. Beforethegageisused,thechambertemperature, inkelvins,mustbe enteredfromthefrontpanelofthecontrollerasaparameter. Temperature correctionsmustbema...
Open at page →
Page 65
APPENDIXC DIMENSIONSANDCONDUCTANCEOFTHEORIFICE INTHENISTDYNAMICEXPANDER* Introduction TheorificeintheNISTdynamicexpansionapparatuswasdesignedand fabricatedin1980tosatisfythreeconditions; 1. Sufficientthicknesstoensuredim...
Open at page →
Page 66
Gallagher,whoreportedavalueItCq=0.4425in±0.00003in,whererQisthe radius. Eachofthesphericalsegmentsterminatesatanaxialdistance b=0.0150in±0.0005infromtheplaneofthethroat. Transitiontothefull thicknessoftheorificeplateisma...
Open at page →
Page 67
0.99995< <0.99997. Acarefullydesignedconicaltransitioninplaceofanabruptstepwould requireacorrectionfactormuchclosertounity. Unfortunately, designand fabricationprecededrecognitionofthispoint. Fromthecomputedresultsandunc...
Open at page →
Page 68
-W= J h(z)p(z)dz, (C2) -b wherep(z),translating itsverbaldefinitionintomathematical form,isthe solutionoftheinhomogeneous Fredholmintegralequationofthesecondkind p(z)= Po(z) + [ g(z,u)p(u)du. (C3) J-b Thesetwoequationspr...
Open at page →
Page 69
Becauseofeq(C6)itsuffices,toexpandtheintegraleq(C3)onlyinthe rangeofpositivez.Aftersomemanipulation theresultbecomes pb 2Rp(z)=R-a+A-2R g(-z,u)p(u)du; 0<z<b , where j.b p(u)du. Integratingeq(C8) where (2R-b)A=b(R-a) -B,...
Open at page →
Page 70
Gas Conductance CL/s) H2 43.52 He 30.89 N2 11.675 Ar 9.777 Theeffectofdeviationsfromthecosinelawofscatteringisbeyondthe scopeofthisreport.Asimplemodel,basedontheassumptionthatasmall fractionofmoleculesundergoesspecularre...
Open at page →
Page 71
APPENDIXC-I ListofSymbols R: radiusofcurvatureofsurfacesAandB(fig.7). r^, :radiusofthethroat , a: distancefromthecenterofthethroattothecenterofcurvatureof surfaceAorsurfaceB. a^= -r^,^ b: distancefromthecenteroftheentran...
Open at page →
Page 73
APPENDIXC-II TheIntegralAandtheTransmissionProbabilityK TheIntegralAandthetransmissionprobabilityKhavebeennarrowly boundedwithoutsolvingintegraleq(C8) .Firsttheintegrals f ^ I(z)=2R g(z,u)du; -b<x<0 J 0 and J= I(z)dz J-b...
Open at page →
Page 74
Assigningnominalvaluestotheorificedimensionsandsolvingthissystemfor thebounds, 0Ti476T<~i(zyT'O48l27^^^^^^ 0<z<b Fromeq(C12)andtheseboundsonA, ' 0.991672<t<(T.991677. HenceK=0.99167tofivedecimalplaces,withanuncertaintyof...
Open at page →
Page 75
NBS-n4A(REV.2-80) U.S.DEPT.OFCOMM.. BIBLIOGRAPHICDATA SHEET(Seeinstructions) 1.PUBLICATIONOR REPORTNO. NIST/SP-250/34 2.PerformingOrgjui.ReportNo,3.PublicationDate March1989 4.TITLEANDSUBTITLE NISTMeasurementServices:Hig...
Open at page →
Page 79
THESP250SERIESONNISTMEASUREMENTSERVICES* SP250-1 Spectra)RadianceCalibrations SN003-003-O2792-8 $3.50 SP250-2 FarUttravloletDetectorStandards PB87227609 SP250-3 RadiometricStandardsintheVacuum Uitraviolet PB87227625 SP25...
Open at page →
Page 80
U.S.DepartmentofCommerce NationalInstituteofStandardsandTechnology (formerlyNationalBureauofStandards) Galthersburg.MD20899 OfficialBusiness PenaltyforPrivateUse$300
Open at page →