Page 1
NBSMeasurementServices: Regular Spectral Transmittance NBS Special Publication 250-6 U.S.DepartmentofCommerce NationalBureauofStandards
Open at page →
Page 2
CenterforRadiationResearch TheCenterforRadiationResearchisamajorcomponentoftheNationalMeasurementLaboratory intheNational BureauofStandards. TheCenterprovidestheNationwithstandardsandmeasurementservicesforionizing radiat...
Open at page →
Page 3
NBSMEASUREMENTSERVICES: REGULARSPECTRALTRANSMITTANCE KennethL.Eckerle JackJ.Hsia KlausD.Mielenz VictorR.Weidner CenterforRadiationResearch NationalMeasurementLaboratory NationalBureauofStandards Gaithersburg,MD20899 U.S....
Open at page →
Page 4
LibraryofCongressCatalogCardNumber:87-619840 NationalBureauofStandardsSpecialPublication250-6 Natl.Bur.Stand.(U.S.),Spec.Publ.250-6,59pages(July1987) CODEN:XNBSAV Commercialproducts—materialsandinstruments—areidentifiedi...
Open at page →
Page 5
PREFACE Thecalibration andrelatedmeasurement servicesoftheNationalBureau ofStandardsareintendedtoassistthemakersandusersofprecision measuringinstruments inachievingthehighestpossiblelevelsof accuracy,quality,andproductiv...
Open at page →
Page 6
Abstract Thisdocumentation describesmeasurement services,instrumentation, standards, andmeasurementtechniquesforregularspectraltransmittance over thespectralrangeof200to2500nm. Itisdividedintothreeparts. Part Idescribest...
Open at page →
Page 7
TABLEOFCONTENTS Paqe Abstract iv I.Measurement Services,PhilosophyandAvailableStandards 1 A.MeasurementPhilosophy 1 B.Description ofServices 1 C.StandardReferenceMaterials andMeasurementAssuranceProgram 3 II.ReferenceSpe...
Open at page →
Page 8
TABLEOFCONTENTS(cont'd) Page III.TransferSpectrophotometer forRegularTransmittance 35 A.DescriptonoftheInstrument 35 1.WavelengthRange 35 2.BandwidthRange 35 3.PhotometricRange 35 4.TemperatureControl 36 5.DataAcquisitio...
Open at page →
Page 9
TABLEOFCONTENTS(cont'd) Page LISTOFTABLES 1.CalibrationServicesforSpectrophotometry 2 2.StandardReferenceMaterialsforSpectrophotometry 4 3.Range,Source,Grating,DetectorandMeasurement Modefor WavelengthRanges 18 4.Wavelen...
Open at page →
Page 10
LISTOFFIGURES Page Fig.1.DiagramPertainingtotheDefinitionofRegular Transmittance 6 Fig.2.SchematicofReferenceSpectrophotometer asUsedinthe Ultraviolet andVisible 10 Fig.3.SchematicofReferenceSpectrophotometer asUsedinthe...
Open at page →
Page 11
I.Measurement Services,Philosophy andAvailableStandards A.MeasurementPhilosophy Regularspectraltransmittance measurements provideusefulinformation onthepropertiesofmanymaterials. NBSstrivestoprovidethebest possiblestanda...
Open at page →
Page 12
Table 1 Calibration ServicesforSpectrophotometry Old New Instrument Number Number 7.8A 38060S R,T Spectrophotometric measurementservices (SpecialTestsonly)Transmittance and Reflectance. 7.8B 38010C T Spectraltransmittanc...
Open at page →
Page 13
Withthetransferinstrument, surveyspectralscansareavailableona limitedbasisforapplications wheresuchsurveysofsamplesareneededbut highaccuracy isnotrequired. Forareducedfee,uncorrecteddigitalor analogdataareprovidedwithout...
Open at page →
Page 14
Table2 StandardReferenceMaterialsforSpectrophotometry Special PublicationSRMNumber Description 260-51 930 GlassFiltersforSpectrophotometry 260-51 2030 GlassFilterforTransmittanceMeasurements 260-68 2031 Metal-on-Quartz F...
Open at page →
Page 15
Transmission isthepassage(withoutchangeinwavelength) ofoptical radiationthroughtran-sparent ortranslucentmedia. Transmissionmaybe regular(withoutchangeindirection) ordiffuse(spreadingoutindifferent directions). Transmiss...
Open at page →
Page 16
6
Open at page →
Page 17
andthesame. Letn=n(l+ik)bethecomplexindexofrefraction ofthe slabmedium, n'theindexofrefraction oftheinitialandfinalmedia,and 9'theangleofincidence. Undertheseconditions, theregularspectral reflectance oftheslabisgivenby:...
Open at page →
Page 18
rapidlyoscillatinginterferencetermdependingon(t/x)inadditiontothe attenuationtermthatdeterminestheabsorptionofthefilterasafunction oftheattenuationcoefficient kandfilterthickness t.Asaconsequence, thetransmittance T(x,0'...
Open at page →
Page 19
B.Instrumentation TheNBSreferencetransmittance spectrophotometer (RTS)hasbeen described indetailinpreviouspublications [1,2,3,4]. Abriefdescription oftheinstrument, practicalsuggestions foritsuse,andsomeofitslimi- tation...
Open at page →
Page 20
10
Open at page →
Page 21
grooves/mmblazedat1250nmrespectively. Themonochrornator collimatinq andfocusingmirrorsareoff-axisparabolas,paraboloidal towithin0.1 fringeingreenlightandfreeofzones. Thesemirrorscanbemovedalonga railandadjustedbymicromet...
Open at page →
Page 22
i5 Q.z2 Oq:<E XUJ^ OO Q. UTER 2>a: 4 — IRENT EQUEN NVERT OUNTE 'Willr o 50:00 -a s- S- rO O) -a 0) (/) in 03 S- cu 4-> (U o -l-> o O- o s- +J o cu Q. iA CU o (U s- ^- OJ s- <4- O a •r— 4-> rd E CU o C7) 12
Open at page →
Page 23
lock-inamplifierwherevoltageoutputisconvertedtoacurrent. Otherwise, theinstrument isoperated inthesamemannerasintheultraviolet and visible. 2.OpticalAlignment TheopticalalignmentoftheRTSisbasedonpre-determinedvaluesof th...
Open at page →
Page 24
4.MeasurementCapabilities andRequiredSampleProperties Thefollowinginformation isprovidedtoacquaintuserswiththetypesof measurements thatcanbeperformedontheRTS.Alsolistedaretherequired propertiesofuser-submitted samplestha...
Open at page →
Page 25
takentoavoidcontactwiththemetalfilm,sincescratchesandothermarks cangreatlydegradesuchafilter. d.MeasuredArea. Theusualprocedureforthevisiblespectralregionistomeasurean averagetransmittance overacentralareawhichissmallert...
Open at page →
Page 26
polarizer inthespectrophotometer beamalongwiththefilter,andmeasure thetransmittance ofthecombinationwiththefilterrotated in45° incrementsaboutitsnormalwhilethepolarizerremainsfixed,orvice-versa. Forauniformnon-polarizing...
Open at page →
Page 27
j .FiIterSize. FiltershavingthefollowingcommonsizescanbemeasuredontheRTS eithersinglyoringroupsoftwoorthree: 5.1cmX5.1cmsquare 3cmdiametercircular 1.3cmX1.3cmx5.7cmcuvette 1.1cmX3.05cmrectangle. (Othersizesoffiltersupto1...
Open at page →
Page 28
Table3 Range,Source,Grating,DetectorandMeasurementModeforWavelengthRanges Wavelength Source Grating Detector Niode (nm) 200-400 DeuteriumArc 600l/mm,200nm PhotomuItiplier D.C. (28w) blaze 360-800 TungstenStripLamp 600l/m...
Open at page →
Page 29
levelsofuncertaintythanthatformeasurementsmadewiththeRTS.) Unless thereisadefiniteneedforaqivenbandwidth, itisusuallybesttochoose thelargestbandwidthwhichcanprovidetheinformation aboutthefilter. Forfilterswithstrongspect...
Open at page →
Page 30
TABLE4 Wavelengths ofAtomicEmissionLinesUsedforWavelength Calibrations intheUVandVisibleSpectrum a.Linesusedwiththe.25mmapertures L. 1CMIC 11I' rtuVC 1CI1U1 1 F1pmpn 1" wavc 1ci 1^L1 1 1nm 1 vnm ; *TJ •yj0\J Hpnc JOO•OUJ...
Open at page →
Page 31
Table5 Wavelengths ofAtomicEmissionLinesandMolecularAbsorptionBands UsedforWavelengthCalibrations intheNearIRSpectrum ElementorChemical (nm) TypeofSpectra Rb 780.02 EmissionLine Rb 794.76 EmissionLine Hg 1013.98 Emission...
Open at page →
Page 32
(uiu)N0ii3ad»03 SAiiiaav 22
Open at page →
Page 33
Theseuncertainties areupperboundsandmayberegardedashaving a99.7% confidence level(3aequivalent). Allupperboundsinthispublication will beequivalentto3auncertainties. 2.LinearityofthePhotometric Scale, Allmeasurements with...
Open at page →
Page 34
.2 .4 .6 .8 1.0 Fig.5:Additivecorrection,aT,versustransmittance T(<^) Theshadedarearepresents±1standarddeviation. 24
Open at page →
Page 35
3.StabilityofSourceandDetectionSystem ThestabilityrequiredoftheRTSissuchthatafilterwithtrans- mittance0.5canbemeasuredwithastandarddeviationof0.00005. Two sourcesofinstabilityhavebeenfoundtodate. Oneisthephotomult i piie...
Open at page →
Page 36
2.SampleAlignment. Thealignmentlaserisusedtoaidinpositioningthesamplesbymeansof threealignmentscrewsonthesamplewheelorslider. Withthewheel, samplesareusuallyalignednormaltothebeam,sothatthelaserlightis reflectedbackintot...
Open at page →
Page 37
Thedarkreadingsareinterpolated intimeandsubtractedfromeachclear- spaceorfilterreadingbeforeanyfurtheranalysis, ^01 to5 Let t^ = 2 (6) bethetimehalf-waybetweentimestgiandto5atwhichreadings IqiandIqs aretaken. Toreducethen...
Open at page →
Page 38
andanew iscalculatedusingEq,(9b). Thisprocess isrepeateduntil s2issufficiently small(lO"^^- TheI^-'sgivingthiss^areusedto calculatethetransmittance forthenthfilter. T=— , n=1,2,3 Thesamplestandarddeviation s^of is ^n = "...
Open at page →
Page 39
andwhereI(vt)istheirexpectation value,obtainedfromaleast-squares fit ofthemeasureddataasalinearfunctionoftime. Thevalvesoperatingthetwoapertureshuttersareactuatedbythedata acquisition andcontrolsystemofthespectrophotomet...
Open at page →
Page 40
wheredlisthephotometricerrorofthespectrophotometer definedbyEq. {15a). E.Uncertainties 1.Instrumental Bias BecauseofthespecialdesignoftheRTS,thecombinedsystematicerror introducedbythespectrophotometer itselfintomeasureme...
Open at page →
Page 41
featuresaremeasured,theerrorcalculatedfromtheaboveformulawillbe included intherandomerrorquotedforthemeasurement. Straylight: Theerrorduetostraylightalsodependsonthe transmittance spectrumofthesample. Fromthemeasurements...
Open at page →
Page 42
32
Open at page →
Page 43
valuesofnon-uniformity are0.1to0.5percentofthetransmittance (equivalent toa3asystematicuncertainty). Thermochromism becomes aproblemforthemeasurement ofcoloredfilters. Whensuchfiltersaremeasured, itisnecessarytomonitorth...
Open at page →
Page 44
Table6 TypicalOver-allUncertainties forNBSReferenceSpectrophotometer Transmittance SourceofError T = 0.01 T=0.1 T = 1.0 A.InstrumentalBias(Systematic) 2.0X10-5 1.0X 10-'+ 1.0X 10-'+ B.RandomError \oXoLanaaruueviaLion^ 1....
Open at page →
Page 45
loweringtheenclosuretoensurethatturningtheliftmechanismresults in movementoftheenclosure. Theusualprecautions areobservedwhenusing chemicalsforstraylighttests. Otherprecautionstakenaretheusualones whenworkingwithelectric...
Open at page →
Page 46
parameterdisplaymonitorscreen,whenthedensityexceeds4.5.Thissame message isdisplayed inthetransmittance modewhenthetransmittance isless than0.01%. Inordertomeasurehigherdensitiesthan4.5,specialphoto- etchednickelscreensar...
Open at page →
Page 47
Table7 EmissionLinesandAbsorptionBandsUsedtoCalibrate theTransferSpectrophotometer WavelengthScale. EmissionLines: Hg:253.65nm,334.15nm,435.83nm,1014nm,1529nm. D2:486.0nm,656.10nm,1312.2nm,1968.3nm,2624.4nm. Note: Thelas...
Open at page →
Page 48
TheD2linesareproducedbythedeuterium lampbuiltintothespectro- photometer, andareusedinautomatedday-to-daywavelengthscalechecks. In theUVandvisiblethe486.0nmand656.1nm linesareusedinthefirst gratingorder,andinthenearIRthey...
Open at page →
Page 49
Thenon-linearity oftheleadsulfidedetectorwasdeterminedby comparingtransmittance measurements at770nm.Thefollowingdifferences werefound,andareappliedasnon-linearitycorrections formeasurements madewiththeleadsulfidedetecto...
Open at page →
Page 50
4.Stability Thewavelengthscaleerrorwillremainconstantafterawarm-uptimeof approximately onehour,ifthetemperature oftheinstrumentremainsstable. Thephotometricscaledriftischeckedbyobservingchangesinthebaseline (100%)calibra...
Open at page →
Page 51
2.MeasurementSequenceforTransmittance orAbsorbance Themeasurements arerecorded inthesequence: 100%,testsample,100%. Intheabsorbancemode,thesequence isthesameexceptthatthe100%becomes thezeroabsorbancecalibration. 3.DataRe...
Open at page →
Page 52
used,theuncertaintyduetodriftwillnotexceed0.0015. Thedriftis linearandcanbeminimizedbyrecordingthe100%calibration atshortertime intervals andcorrectingforchangesinthiscalibration. Theestimated typicaluncertainties (3o)ar...
Open at page →
Page 53
Table8 SummaryofEstimatedUncertainties inPercentofSpectral Transmittance ValuesforNBSTransferSpectrophotometer [A Representative Analysis(ThreeSigma)intheVisibleRegionfor TransmittancesintheRange.1-1.0] Wavelength Strayr...
Open at page →
Page 54
REFERENCES 1.Mielenz, K.D.,Eckerle, K.L.,Madden,R.P.,andReader,J.,"New ReferenceSpectrophotometer", Appl.Opt.j2(7),1630(July1973). 2.Mielenz, K.D.,andEckerle, K.L.,"Design,Construction, andTesting ofaNewHighAccuracySpect...
Open at page →
Page 55
References(cont'd) 11.CIEPublication No.17(E-1.1)1970,International LightingVocabulary , 3rdEdition,BureauCentraldelaCIE,Paris(1970). 12.Eckerle, K.L.,Venable, W.H.Jr.,andWeidner, V.R.,"Averaging SphereforUltraviolet, Vi...
Open at page →
Page 57
APPENDIXA U.S.DepartmentofCommerce NAT[ONALBUREAUOFSTANDARDS GAITHERSBURG.MARYLAND20899 REPORTOFTEST for SpectralTransmittance of FourNeutralFilters Submittedby LaboratoryA (Authorization forthistestwasgiveninPurchaseOrd...
Open at page →
Page 58
APPENDIXA REPORTOFTEST Page2 withtheelectricvectoroftheincidentradiationhorizontal (0°)and vertical(90°). Thesolidanglesubtendedatthesamplebythedetec- tionsystemwasapproximately 0.0027steradians. Theilluminatedarea onthe...
Open at page →
Page 59
APPENDIXA Erroranalysisformeasurements oftransmittance usingthereference Thetransmittance Tofafilterrelativetoaircanbecalculated fromtheexpression whereT„isthetransmittance ofareferencefilter,andTnisthetrans- mittanceoft...
Open at page →
Page 60
X I—I <u Q r— LU fO a. h- a. to CO ^s cy»0CM LlJ CO C\J 10 >0000 0000 0000 0000 0000 4J +-> +-> < • • • • 0000 4-) O) to (O S- •t-> •r" (O UJ O ^ ^ to CM to0 •r— 0 0)0 0toCM u CTt^ ^s CM CX)to0 toc—UJ LU CD r— 1 — CO fO...
Open at page →
Page 61
-o o s- CU O «3 c -o +->o 1 — • Stan of iati E s_ -a £Z to . O) 03 +-> +J S- 000toLO o 'r—c 00«^CO0 CO (U 4- CO •sd-CMCMCM • tooo 000000 •I — 000000 CU CU 00000 vo ,— ^ o 00000 O -M i- 1—00000 4->O <00000 fO +->o LU fO t...
Open at page →
Page 63
APPENDIX B U.S.DepartmentofCommerce NATIONALBUREAUOFSTANDARDS GAITHERSBURG.MARYLAND20899 REPORTOFTEST for SpectralTransmittance of OneTiffenNeutralDensityFilter Submittedby: Laboratory B (ReferencePurchaseorderNo.ABCdate...
Open at page →
Page 64
APPENDIXB REPORTOFTEST Page2 Thedataforthetestfilterwerecorrectedforsmallvariations inthezerocalibrationandconvertedtoequivalentvaluesofspectral transmittance. Theresultsofthedatareductionarelistedintable1. 5.Uncertainti...
Open at page →
Page 65
APPENDIXB TABLE 1 TIFFENN.D.FILTER60%#511 TRANSMITTANCE WAVELENGTH (NM) {%) 380 59.95 390 61.82 400 62.70 410 62.59 ' 420 62.42 430 62.33 440 62.22 450 62.37 460 62.56 470 62.57 480 62.57 490 62.56 500 62.60 510 62.60 52...
Open at page →
Page 67
roilMNBS-l14A(REV.11-84) U.S.DEPT.OFCOMM. BIBLIOGRAPHICDATA SHEET(Seeinstructions) | NBS/SP-250/6 1.PUBLICATIONOR REPORTNO. 2.PerformingOrgan.ReportNo 3.PublicationDate July1987 4.TITLEANDSUBTITLE NBSMeasurementServices:...
Open at page →
Page 69
I
Open at page →
Page 71
PUBLICATIONS INTHESP250SERIESFROMTHECENTERFORRADIATIONRESEARCH SP250-1 NBSMeasurement Services: SP250-12 SpectralRadianceCalibrations J.H.Walker,R.D.Saunders, andA.T.Hattenburg SP250-2 NBSMeasurement Services: SP250-13 F...
Open at page →
Page 72
U.S.DepartmentofCommerce NationalBureauofStandards Gaithersburg,MD20899 OfficialBusiness PenaltyforPrivateUse$300
Open at page →