Cw Laser Power and Energy Calibrations
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
NAT'LINSTOFSTAND&TECH A111D7DITDMD NI8T PUBL!CAT!0N8; NISTMeasurementServices: nist cwLaserPowerandEnergy o^uf^'* ^ ,a11^-^ Publication CalibrationsatNIST 250-75 JoshuaHadler Christopher L.Cromer JohnH.Lehman <3C ioo •^5...
Open at page →
Page 2
rhe NationalInstituteofStandardsandTechnologywasestablishedin1988byCongressto"assistindustryinthe developmentoftechnology ...neededtoimproveproductquality,tomodernizemanufacturingprocesses,toensure productreliability ......
Open at page →
Page 3
NISTSpecialPublication250-75 NISTMeasurementServices: cwLaserPowerandEnergyCalibrationsatNIST JoshuaHadler Christopher L.Cromer JohnH.Lehman Optoelectronics Division ElectronicsandElectricalEngineeringLaboratory National...
Open at page →
Page 4
Certaincommercialentities,equipment,ormaterialsmaybeidentifiedinthisdocument inordertodescribe anexperimentalprocedureorconceptadequately.Suchidentification isnotintendedtoimply recommendation orendorsementbytheNationalI...
Open at page →
Page 5
Contents Page 1. Introduction 1 2. CalibrationServiceSummary 1 3. TheoryofMeasurement 2 3.1 SystemBasicsandMeasurementPrinciple 2 3.2 MeasurementEquations 3 3.3 DataAnalysis 4 4. MeasurementSystem 7 4.1 MeasurementStanda...
Open at page →
Page 7
cwLaserPowerandEn&rgyCalibrationsatNIST JoshuaA.Hadler,ChristopherL.Cromer,andJohnH.Lehman NationalInstituteofStandardsandTechnology 325Broadway Boulder,Colorado80305 Thisdocumentdescribesacalibrationserviceforabsolutela...
Open at page →
Page 8
deviceundertest(DUT),issenttoNIST,whereitisthencomparedtotheappropriatelaboratory standardusingacahbratedbeamspUttermeasurementsystem. TheDUTmaysimplybeastandalonedetector,oritmaybeanintegratedsystemwithan independentdis...
Open at page →
Page 9
Twocalibrateddetectors(standards)areusedtocalibratetheDUTbysubstitutionoftheDUT foroneofthestandards.Awedgebeamsplitter ispositionedtodirectareflectedbeamtoone standardandthetransmittedbeamtoasecondstandard.Theopticalpro...
Open at page →
Page 10
(3.2) ThevalueofthebeamspHtterratiocanbedeterminedbysimultaneouslymeasuringboththe transmittedandreflectedbeamsincidentonthestandarddetectors.Thisgivesusthebeamsplitter ratioRintermsofthelaserpowermeasuredbyeachstandard....
Open at page →
Page 11
ashutter-openperiod,duringwhichthelaserimpingesontheDUT.Laserpowermeters generallyhaveanoutputthatreachesanequilibriumvaluesometimeaftertheshutterisopened, whichiscalledthesettlingtime. Thedetector'ssetthngtimeisderivedf...
Open at page →
Page 12
integratedfromthedataacquiredduringtheperioddefinedbytheopeningoftheshuttertothe endofthesetthngtimeaftertheshutterisclosed.Othertypesofenergymetersmayrequirea measurementofthepeakoutputfromtheDUTastheoutputtouseinthecaH...
Open at page →
Page 13
4.MeasurementSystem 4.1MeasurementStandards Energyandpowermeasurementscanbeacquiredusingavarietyoftypesofdetectors.Such detectorsmightinclude,butarenotHmitedto,thermopiles(surfaceandvolumeabsorbers), calorimeters,anddiod...
Open at page →
Page 14
4.3OpticalSetup Standard1 Tratulatian Stage 12 Eiclosijifi Slmtter Attenualnr EntranceIris ML'> Beamsplitter EnclosijM Figure4.3.Exampleopticallayout. ThebasicopticalsetupisrepresentedhereinFig.4.3. 4.3.1Attenuator Anatt...
Open at page →
Page 15
measurementlabmayexperience.Eachenclosurecontainsitsownprecisionmeasurement electronicstoavoidunnecessarilylongsignalpathlengthstothedataacquisitionelectronics. Thesemetersarethenreadremotelyviacomputerintothedataacquisi...
Open at page →
Page 16
5.StandardOperatingProcedures 5.7SystemStartup Allelectronicsandothersupportsystemsarepoweredupatleastonehourpriortothestartofthe measurement.Alltestequipmentisallowedtoequilibratewithroomtemperature.Thelaseris selectedf...
Open at page →
Page 17
6.Setadjustableaperturesinthebeampathstoblockscatteredlightandstraybeams,but notsoastooccludethemeasuredbeam. 5.3DUTAlignment TheDUTismountedonatranslationstagesuchthatitcanbeeasilyinsertedandremovedfrom theincidentbeam....
Open at page →
Page 18
5.Standard/monitor gain. 6.Background,orratingperiod. 7.Injectiontime. 8.Settlingtime(usually7xtimeconstant).Thesettlingtimeisgenerallyestablishedbythe manufacturer's specification.Wherethesettlingtimeislessthanthatofthe...
Open at page →
Page 19
6.UncertaintyEvaluation TheuncertaintyestimatesfortheNISTlaserpowerandenergymeasurementsareassessed followingguidelinesgiveninNISTTechnicalNote1297,"GuidelinesforEvaluatingand ExpressingtheUncertaintyofNISTMeasurementRes...
Open at page →
Page 20
6.1.1SpectralResponsivity(TypeB) Spectralresponsivity isgenerallydefinedasoutputperincidentpowerand/orenergy.Itis expressedmostoftenasresponsivityataspecificwavelength,mostnotablyfordiode-based detectors,whichcanhavelarg...
Open at page →
Page 21
thedetectorresponseasafunctionoftemperature.Anexampleofsuchbehaviorisdocumented inNISTSpecialPublication250-53[19]. 6.1.6WindowTransmittance(TypeB) Foradetectorwhereresponsivity isdeterminedbymethodsotherthandirectsubsti...
Open at page →
Page 22
6.2MeasurementSystemUncertainties Theuncertaintiesforthisserviceareincludedineachcalibrationreport,andaresummarized here.Theerrorbudgetissummarizedbelowintable6.2. 6.2.1 InjectionTime(TypeB) Theevaluationofenergymetersan...
Open at page →
Page 23
6.2.6TransferMeasurement(TypeA) Theaccuracyofthedirectsubstitutionmeasurement isquantifiedbythestandarddeviationof thetransfer-standard cahbrationfactors(seeEqs.3.4aand3.4b). Table6.2.Examplesummaryofsystemuncertaintie.y...
Open at page →
Page 24
7.QualityControl TheOptoelectronicDivisionmeasurementservicesmakeuseofqualityassurancepracticesto ensurethevalidityofmeasurementresultsandtheiruncertainties.Suchpracticesinclude: •Repeatedmeasurements/calibrations compar...
Open at page →
Page 25
9.Glossary Calorimetry: Themeasurementofthequantityofheatexchanged.Inthisservice,thecalorimeter isusedto quantifytheamountofthermalenergydepositedbyalasersource. DirectSubstitution: SubstitutingaknownstandardforaDUTinide...
Open at page →
Page 26
Material;FusedSilica AngleofIncidence:-8,76° Scale: 6mm Figure9.1.Diagramofwedgebeamsplitter. StandardUncertainty: Therepresentationofeachcomponentcontributortotheuncertaintyofameasurement result. Typicallyexpressedwitht...
Open at page →
Page 27
10.References [I]BarryN.Taylor,TheInternationalSystemofUnits(SI).NISTSpecialPublication330;2001. [2]BarryN.Taylor,GuidefortheuseoftheInternationalSystemofUnits(SI).NISTSpecial Publication811;1995. [3]XiaoyuLi,ThomasScott...
Open at page →
Page 28
AppendixA.MeasurementSystemUncertainties A.lInjectionTime(TypeB) Anupperboundwasestablishedbasedontheshuttermanufacturer's specifications,whichare thefollowing: Shutter:Windowspeed=48ms ShutterController:±10%ofshutterspe...
Open at page →
Page 29
A.3LaserPowerDrift Noisethatistermeddrift(or\lfnoise)iscommontoabroadrangeofphysicalsystemsandis observedatlowfrequencies(havingamagnitudeinverselyproportionaltofrequency). Itis typicallynotbasedonasinglephenomenon;rathe...
Open at page →
Page 30
Modeled3%LaserPowerDriftOver200sInjectionPeriod FigureAl.Laserpowerdriftmodel. Reference [Al]D.W.Allan,Shouldtheclassicalvariancebeusedasabasicmeasureinstandardstechnology?IEEETrans. Instrum.Meas.IM-36:646-654;1987 24
Open at page →
Page 31
AppendixB.SampleCalibrationReport U.S.DEPARTMENTOFCOMMERCE NATIONALINSTITUTEOFSTANDARDSAND TECHNOLOGY ELECTRONICSANDELECTRICALENGINEERING LABORATORY Boulder,Colorado80305 ReportofCalibration(42110CA) LASERPOWERMETER <DUT...
Open at page →
Page 32
LASERPOWERMETER DetectorModelNo: DisplayModelNo: Figure1.Measurementsetup. UncertaintyAssessment TheuncertaintyestimatesfortheNISTlaserenergymeasurementsareassessedfollowingguidelines giveninNISTTechnicalNote1297,"Guidel...
Open at page →
Page 33
LASERPOWERMETER DetectorModelNo: DisplayModelNo: Sr = (Zx)^ N N-1 wherethexvaluesrepresenttheindividualmeasurementsandNisthenumberofxvaluesusedfora particularTypeAerrorcomponent.ThestandarddeviationofthemeanisSr/N'^,andt...
Open at page →
Page 34
LASERPOWERMETER DetectorModelNo: DisplayModelNo: FortheDirector, NationalInstituteofStandardsandTechnology CalibratedBy: MariaL.Dowell,GroupLeader Sources,DetectorsandDisplaysGroup OptoelectronicsDivision JoshuaHadler,Ph...
Open at page →