Laser Beam Diameter Instruments
Inside this document
Every page is full-text indexed — open a page directly in the PDF viewer.
Page 1
NIST PUBUCAJJONil REFEflEMCE NISTSpecialPublication250-87 CalibrationServiceforInstruments thatMeasureLaserBeamDiameter ShaoYang 1^ll^^^^n NationalInstituteofStandardsandTechnology I^1^HilrI UnitedStatesDepartmentofComme...
Open at page →
Page 2
rhe NationalInstituteofStandardsandTechnologywasestablishedin1988byCongressto"assistindustryinthe developmentoftechnology ...neededtoimproveproductquality,tomodernizemanufacturingprocesses,toensure productreliability ......
Open at page →
Page 3
NISTSpecialPublication250-87 CalibrationServiceforInstruments thatMeasureLaserBeamDiameter ShaoYang Optoelectronics Division NationalInstituteofStandardsandTechnology 325Broadway Boulder,CO80305-3337 July2010 U.S.DEPARTM...
Open at page →
Page 4
Certaincommercialentities,equipment,ormaterialsmaybeindentifiedinthisdocumentinordertodescribean experimentalproceduralorconceptadequatetly.Suchidentification isnotintendedtoimplyrecommentionor endorsementbytheNationalIn...
Open at page →
Page 5
CONTENTS 1. Introduction 1 2. CalibrationServiceSummary 1 3. TheoryofMeasurement 2 3.1 SystemBasicsandMeasurementPrinciple 2 3.2 MeasurementEquations 4 3.3 DataAnalysis 7 4. MeasurementSystem 7 4.1 TestBeam 7 4.2 Scannin...
Open at page →
Page 7
CalibrationServiceforInstrumentsthatMeasureLaserBeam Diameter ShaoYang NationalInstituteofStandardsandtechnology OptoelectronicsDivision 325Broadway Boulder,CO80305 Thisdocumentdescribesthecalibrationserviceforinstrument...
Open at page →
Page 8
4o)beamdiameterdefinedbyISO,butmayusedifferentmeasurementmethodsother thantheNISTmeasurement.Atthecompletionofthecahbrationmeasurements, the DUTandacahbrationreportaresenttothecustomer.Thecahbrationreportsummarizes there...
Open at page →
Page 9
Currentsource &TEcontroller FC/PC connectors Monitor detector LD LDmount 0 Variablefiber attenuator 0 Fibersplitter B. FC/PC connectors Testbeam Collimator Detectionplane Slit&detector NISTbeamdiameter measurementsystem...
Open at page →
Page 10
2 J-ooJ-cO OO (Ty= \ \ • -3.2 3.2MeasurementEquations ISOdefineslaserbeamdiameterbyuseofthesecondordermomentsofthebeampower densitydistribution I{x, [1 - 3]: rr{x-xfI{x,y)dxdy _2 J-ODJ-ao 1 CT= , 3.1 I{x,y)dxdy J-aoJ-QO...
Open at page →
Page 11
f xP{x)dx x=— , 3.10 rP{x)dx J-00 y=— . 3.11 f Q^y)dy J-oO Equations3.5through3.11arethebasisofslitscanmeasurementofthe4f7laserbeam diameter.TheintegrationlimitsinEquations3.1to3.4and3.6to3.11arefromnegative topositivein...
Open at page →
Page 12
3 WidthofIntegrationAreaWg{mm) Figure3.2Anintegrationareawithawidthofthreetimesthebeamdiameterdefinesmeasuredbeam diameterforthetestbeamandthebestfitGaussianbeam.Thesolidanddottedcurvesshowthemeasured beamdiameter, Df,,o...
Open at page →
Page 13
3.3DataAnalysis IntheNISTslitscanmeasurementsystem,thewidthoftheshtis10|imandthelengthis 7mm.Foronesinglecompletemeasurement,theslitscansinboththehorizontal(x)and vertical(y)directionsatstepsof25[imforarangeof8mm.Usingit...
Open at page →
Page 14
changeof0.002%inthefibermodefielddiameter,orachangeof0.05[iminthetest beamwidth. 4.2ScanningSlitMeasurementSystem Forthecharacterizationofthetestbeamdiameter,theNISTmeasurementsystemuses thescanningslitmethod.Thewidthoft...
Open at page →
Page 15
specificsoftheDUT.Eachmeasurementrunmayalsoconsistofseveralmeasurements. EachNISTmeasurementincludesonescanineachofthexandydirections.Thescan takes25|imineachstepandcoversarangeof8mm.Eachscandatumisacquired simultaneousl...
Open at page →
Page 16
wherethexvaluesrepresenttheindividualmeasurementsandNisthenumberofxvalues usedforaparticularTypeAerrorcomponent.Thestandarddeviationofthemeanis Sr/N'^,andthetotalstandarduncertaintyofthemean is[I(Sr^/N)] wherethe summati...
Open at page →
Page 17
themonitordetector.Simulationresultsindicatedthatthisresultedinanuncertaintyof 0.1|iminthebeamwidthmeasurement. 6.1.2InterferencePattern(TypeA) Becauseofthenarrowspectralwidthofthelaserbeam,interferenceduetoreflections b...
Open at page →
Page 18
6.2.5DetectorNonlinearity(TypeB) NonlinearitymeasurementoftheNISTdetectorshowedapositiveresponsenonHnearity ofupto2%occurredatthelowerendofthemeasurementrange.Simulationsindicated thatthisresultsinastandarduncertaintyof0...
Open at page →
Page 19
7.QualityControl ThemeasurementservicesoftheOptoelectronicDivisionmakeuseofqualityassurance practicestoensurethevalidityofmeasurement resultsandtheiruncertainties.Such practicesinclude: •Repeatedmeasurements/calibrations...
Open at page →
Page 20
service.Whenavailable,historicdatafrompreviousmeasurementsofthetestbeamshall beplacedintothetestfolderbytheMeasurement ServicesCoordinator afterthe preparationofthecalibrationreport.TheCalibrationLeaderandtheGroupLeaders...
Open at page →
Page 21
References [1] ISO11146-1:2005,''Lasersandlaser-relatedequipment-Testmethodsforlaser beamwidths,divergenceanglesandbeampropagationratios-Part1:Stigmatic andsimpleastigmaticbeams,"2005. [2] ISO11146-2:2005,"Lasersandlaser...
Open at page →