Standards profile: OIML R 111 and what class E2 really promises
When a weight is sold as 'class E2', OIML R 111 is the document making the promise — and the promise is more than a tolerance. The recommendation specifies maximum permissible errors by class and nominal value, but also density limits, magnetic properties, surface condition, and marking, because a weight that meets its MPE while being magnetic or porous will still ruin precision weighings.
The class ladder (E1, E2, F1, F2, M1…M3) is designed around dissemination: each class is intended to calibrate the one below it with adequate margin, from national standards down to the M-class weights that verify shop scales under OIML R 76's non-automatic weighing instrument rules.
The library's SOP 34 shows one of the less obvious corners of practice: selecting sensitivity and tare weights whose own errors don't quietly consume the budget of the weighing they support. R 111's tables plus that procedure are most of what a working mass bench needs daily.
The class ladder (E1, E2, F1, F2, M1…M3) is designed around dissemination: each class is intended to calibrate the one below it with adequate margin, from national standards down to the M-class weights that verify shop scales under OIML R 76's non-automatic weighing instrument rules.
The library's SOP 34 shows one of the less obvious corners of practice: selecting sensitivity and tare weights whose own errors don't quietly consume the budget of the weighing they support. R 111's tables plus that procedure are most of what a working mass bench needs daily.
Sources — reference chain
Every claim traces to an authoritative reference; Atlas standards entries link onward to the official publisher.
OIML R 111-1:2004
Weights of classes E1, E2, F1, F2, M1, M1–2, M2, M2–3 and M3 — Part 1: Metrological and technical requirements — Atlas entry → official source at www.oiml.org
ATLAS INDEX
OIML R 76-1:2006
Non-automatic weighing instruments — Part 1: Metrological and technical requirements — Tests — Atlas entry → official source at www.oiml.org
ATLAS INDEX
SOP 34 — Selection Use Sensitivity Weights Tare Weights
Atlas library — full text, page-searchable (5 pages, Jun 2019)
ATLAS LIBRARY
More news
Inside GMP 11: setting calibration intervals you can defend
Aug 8, 2026
Torque explainer: why ISO 6789 splits design from calibration
Aug 7, 2026
Control charts I: SOP 9 and the discipline of watching your process
Aug 6, 2026
Inside SOP 2: air buoyancy, the invisible force in every weighing
Aug 5, 2026